How to use the sterilization bioindicator of Shandong Top Biological Engineering Co.

1, put the bio-indicator into the center of the standard test package or the center of the package of items to be sterilized, to ensure that the indicator is in the correct position.

2. Place the sterilized package or item where the bio-indicator is located in the sterilizer and sterilize it according to the specified temperature and time.

3. After the sterilization process, remove the bioindicator from the sterilizer and hold it vertically in your hand with the cap facing up.

4. Use a specialized tool to crush the ampoule in the tube to ensure that the ampoule is broken and the culture solution flows out.

5. Submerge the bio-indicator in the culture solution and incubate at 56℃ or in the matching micro-incubator for 24 hours to observe the preliminary results and 48 hours to observe the final results.