How to debug the DR machine after moving

The steps for debugging a DR machine are as follows:

1. Adjust the position of the DR machine to ensure that the position is stable and that the exposure bed is in the correct position.

2, enter the working state, use the mask to cover the corresponding photocoupler of the cassette, and at the same time, check whether the corresponding LED lights on the cassette control board is normal action (generally cover the limit corresponding to the LED is off, release the mask LED lights up). Be sure to ensure that each limit is working properly before proceeding to the next step in the debugging.

3, motor direction debugging, enter the debugging state, respectively, all the motor direction of the point film box debugging, to ensure that the point film box motor movement direction is accurate. When the motor moves, cover the corresponding limit with the mask to see whether the corresponding motor has action (cover the limit motor stops rotating, and release the mask to rotate the motor). Repeat this process to ensure that each motor and limit of the point film box can work properly.