How to maintain the ultrasonic cleaner?

1. Pay attention to the careful use of cleaning fluid

In the use of ultrasonic cleaning machine, must be in accordance with the specification requirements of the order of operation, in particular, can not be turned on first and then poured into the cleaning fluid, which belongs to the violation of the operation will lead to serious consequences and even damage to the machine, so be sure to put the liquid in the first and then turn on the switch. At the same time should also take into account the composition of the cleaning fluid, if it contains flammable liquids, then be sure to complete the operation in a short period of time, and the whole process should be supervised, not to leave the scene.

2. Pay attention to the use of corrosive or volatile cleaning fluid to do special treatment

Usually do not advocate the use of hazardous cleaning fluids, but if the situation is more special, emphasizing the ultrasonic cleaning machine manufacturers must use corrosive or volatile liquids, then it should be taken indirectly to the proper handling. The specific method of operation is to add water to the cleaning tank first, and then another container of cleaning fluid and put the items need to be cleaned, and then they are immersed in the cleaning tank of water, with this way to clean the will also be in the same effect, but also to ensure the safety of the issue.

3. Ultrasonic cleaning machine power supply and heater power must have a good grounding device.

4. Ultrasonic cleaning machine is strictly prohibited in the absence of water or solvent in the tank, never start, resulting in empty vibration, resulting in vibration head scrapped or damaged.

5. Cleaning equipment with a heating system is strictly prohibited to open the heating switch when there is no liquid.

6. Prohibit the use of heavy objects (iron) to hit the bottom of the cleaning cylinder, so as to avoid damage to the energy converter chip.

7. Ultrasonic generator power supply should be used separately all the way 220V / 50Hz power supply and equipped with 2000W or more voltage regulator.

8. The bottom of the cleaning cylinder should be rinsed regularly, without excessive debris or dirt.

9. Do not put your fingers into the cleaning tank during the operation of the cleaning machine, otherwise you will feel pain or discomfort.

10. Each time you change the new liquid, wait until the ultrasonic start before washing parts.

11. Use water or aqueous solution as the cleaning agent, absolutely prohibit the use of alcohol, gasoline or any flammable gas as a cleaning agent added to the cleaning machine.

12. When the direct use of flammable liquid cleaning, to try to complete the cleaning work in a short period of time, and require the operator not to leave the scene.

13. When the need for corrosive or volatile cleaning fluid, you can use indirect cleaning methods. That is: first of all, add water in the cleaning tank, and then pour the required cleaning fluid into a suitable container and put into the cleaning materials, and then containers containing cleaning fluid and cleaning materials immersed in the cleaning tank, you can start cleaning, the effect is the same as direct cleaning.

14. The best way to make the ultrasonic cleaning effect, ultrasonic cleaning tank temperature is best for 30 ~ 50 ℃, according to the different cleaning objects to correctly select the cleaning agent. Cleaning agent is generally divided into water-based (alkaline) cleaning agent, organic solvent cleaning agent and chemical reaction cleaning agent. Usually use the most water-based cleaning agent, according to the degree of contamination of the cleaned object and the nature of the dirt, choose a different cleaning time.

15. Prohibit the use of the first boot after pouring the cleaning solution, which will cause damage to the machine.

16. Cleaning items should not be placed directly on the bottom of the cleaning tank, the application of special cleaning bracket cleaning baskets, which is conducive to improving the cleaning effect.

17. Try to avoid prolonged continuous work, generally no more than 30 minutes is appropriate.

18. In the use of the tank cleaning or degassing solution do not put too little, generally in the tank 2 / 3 at the best.

19. Pay attention to the factors related to the control of cleaning

Gut ultrasound suggests that we do not need to clean things at the bottom, should be used to specialize in cleaning bracket for the overhead placed in the middle of the position, so that in the feeding time can be uniformly stressed to obtain a good cleaning effect. At the same time, solid as ultrasonic cleaning experts, but also to remind everyone to pay attention to the energy put into the cleaning fluid should be well controlled, can not put too little or may damage the instrument or failure.