Can adipose tissue surface be selected as the adhesive part of negative plate?

It is also possible in theory, but try to stick it in a place with rich muscles. The following are the precautions for pasting the electrode plate:

◆ If there is a metal implant, try to avoid the current between the surgical site and the patch; Therefore, it is necessary to stick an adhesive plate on the surgical current channel that can avoid metal implants.

◆ Be as close to the operation area as possible, and reduce the load resistance of the electrosurgical unit, so that the operation can be carried out with less power. At the same time, the current flow area is reduced, and the possibility of multi-point short circuit or multi-point grounding is reduced, both of which are beneficial to prevent burns.

◆ The neutral electrode plate must be placed in a smooth, hairless, muscle-rich (non-fat), scar-free, and bony protrusion-free position, so as to reduce contact impedance, reduce power consumption, enhance heat dissipation effect, and prevent burns.

◆ Try to choose two electrode pieces. For example, the electrotome produced by Shanghai Hutong has the function of detecting the contact quality of electrode plates. Once the effective contact area is reduced, the electrotome will give an alarm and stop output.

* Choose conductive viscose polar plate, which can fit the skin more effectively and reduce the risk of accidental separation.