DNP-9002 series of electrothermal incubator adopts electric heating to control the temperature inside the working chamber, mirror stainless steel liner, high precision, good insulation performance, widely used in health care, biopharmaceuticals, agricultural research, environmental protection and other research applications. Used for breeding, fermentation, microbial culture, a variety of constant temperature test, environmental testing, material denaturation test and storage of culture media, serum, drugs and other items.
I. Product Technical Parameters
1.1 Temperature range: RT+5℃~65℃;
1.2 Fluctuation: ±1℃;
1.3 Display accuracy: 1℃;
1.4 Workshop size: 350*350*410MM
1.5 Appearance of the approximate dimensions: 472×470×692MM
1.4 Dimensions: 472×470×692mm. 692MM
1.6 Heating speed: ≤2℃/min;
1.7 Power: 150W;
1.8 Electricity: 220V/ 50Hz;
II. Temperature control system and features
2.1 Extremely humanized touch screen, LED clearly display the set temperature, measured temperature, timing time, the parameter display can be switched at will.
2.2 High-responsiveness thermal sensor and environmental scanning microprocessor chip, constituting a high-precision, high-intelligence temperature control characteristics....
2.3 Built-in temperature corrector, can adjust the temperature of the small error. When the working temperature deviates from the set temperature, it can be alarmed automatically
2.4. It has the function of power failure recovery, when the external power supply is restored, the equipment can be automatically according to the original setup program for rapid thermostat. Automatic stop, timer operation, call recovery, parameter memory, parameter password protection, temperature display correction
2.5. Setting timer function, from 1 minute to 9999 minutes of any set working time, the end of the timer that is set to close the display window.
2.6. With modular design of the upper limit of the temperature tracking function, when the measured temperature deviates from the permissible range, the instrument sound (light) alarm and automatically cut off the heating.
Three, box structure and materials
3.1 outside the box material: the outer box is made of high-quality cold plate by phosphoric acid film salt treatment after high-temperature spraying, the color of the blue door and white body
3.2 studio material: mirror stainless steel liner, the use of glass door, which can be convenient for observation of samples, but also the temperature inside the box to minimize the control of changes in temperature. Workshop stainless steel electropolished shelves, shelves can be adjusted with the user's requirements arbitrary height and how many shelves.
3.3 Internal structure: the use of automatic inhibition system to prevent the door from opening the chamber caused by temperature fluctuations. Realize all-round three-dimensional heating effect, to ensure that the working chamber has a good temperature uniformity, unique geometric cavity clean design, to ensure that the sample active perfect protection. The shelves can be adjusted to any height and the number of shelves with the user's requirements.
3.4 insulation material: high-density fiberglass cotton, in the box before and after the right and left up and down the sandwich filled with insulation
3.5 sealing method: the box door and the studio using high-quality door magnetic seals and insulation materials to ensure that the studio of the sealing of the box door for the inner plate of the toughened glass made
Four. Heating system
4.1 special electric heating blanket close to the outer wall of the studio, all-round three-dimensional heating technology to ensure high precision of the temperature, the response of the sensor control system
4.2 heater heating output power uniformly with automatic instrumentation microcomputer algorithmic control
5. Safety Protection System
5.1 Fused protector
5.2 Studio over-temperature protector, sensor failure alarm, the upper limit of the over-temperature alarm
5.3 Independent over-lift preventer, independent over-temperature protector, .
5.4 Fan overcurrent protection
VI. Equipped with technical data and accessories:
6.1 Operation manual
6.2 Product warranty card
6.3 Certificate of conformity
VII. After-sales service:
7.1 Responsible for remote technical guidance to the user
7.2 Timely provision of spare parts and accessories
7.3 Provide technical advice and support during the use of equipment
7.4 Receive customer failure notification within 8 working hours immediately corresponding
VIII. Main components
Intelligent PID control instrument
Silicon controlled
Heat sensor
Special heating blanket