Stainless steel transfer window performance basic requirements

1, laminar flow transfer window cleanliness requirements: Class B;

2, inside and outside the double-layer shell, the internal arc around the processing, to ensure that there are no gaps in the connection;

3, the use of laminar flow design, the airflow direction of the upward and downward mode, the bottom of the 304 stainless steel cold-rolled plate punching design, and set up a reinforcing bar;

4, the filter: the primary filter G4, high efficiency filter H14;

5, wind speed: after the high-efficiency filter out of the wind speed control 0.38-0.57m/s (0.57m/s after high efficiency filter). G4, H14 high efficiency filter;

5, wind speed: after the high efficiency filter out of the wind speed control in the 0.38-0.57m/s (in the high efficiency out of the wind average flow plate below the 150mm test);

6, differential pressure function: display the filter differential pressure (range of high-efficiency 0-500Pa / intermediate 0-250Pa), accuracy of ± 5Pa;

7, control function: fan start/stop button, configure built-in electronic door interlock; set up ultraviolet lamp, design separate switch, when two doors are closed, the ultraviolet lamp should be out of the open state; set up lighting, design separate switch;

8, high-efficiency filters can be disassembled and installed individually from the upper box, which is convenient for maintenance and replacement of filters;

9, set up an access port in the The access port is set in the lower part of the transfer window, which is used for the maintenance of the fan;

10, Noise: Noise <65db during the normal operation of the transfer window;

11, High-efficiency outflow plate: 304 stainless steel mesh plate is used.