(1) Class AP equipment means that under normal use and normal conditions, the equipment, equipment parts or components must not ignite flammable anesthetic gas mixed with air.
(2) Class APG equipment means equipment, equipment parts or components that, in normal use or in a single fault condition, must not ignite flammable anesthetic gases mixed with oxygen or nitrous oxide.