1, ultrasonic cleaning machine ultrasonic power density: the higher the power density, the stronger the cavitation effect, the better the ultrasonic cleaning effect, the faster the cleaning equipment. For precision workpieces such as medical syringes should be used small power density.
2, ultrasonic cleaning equipment, ultrasonic frequency: the lower the frequency, the better the cavitation, the higher the frequency, the better the effect of refraction. For medical syringes should be used for low-frequency ultrasonic cleaning.
3, ultrasonic cleaning machine cleaning temperature: ultrasonic cavitation at 40 ℃ ~ 50 ℃ when the best. The higher the temperature, the more conducive to the decomposition of dirt, but when the temperature reaches 70 ° C to 80 ° C later, it will affect the role of ultrasound, reducing the cleaning effect.
4, cleaning time: in general, the longer the cleaning time of the ultrasonic cleaning machine, the better the results, but for medical syringes in general 3-5 minutes.