(1) Dielectrics Withstanding Voltage
Output Voltage: 0-5kV/AC ±3%
Current Measurement Range: 0-40mA/AC ±3%
Current Alarm Range: 0-40mA within any value ±3%
Arc Detection range: 1-10 levels
Voltage ramp-up time: 1-99s continuously adjustable
Testing time: 1-99s continuously adjustable
Output frequency: 50Hz/60Hz two gears
Drive mode: programmable linear amplifier
(2) Insulation resistance test Insulation resistance test
Test voltage: 500V/DC and 1000V DC two ±5%
Measuring range: 0-200MΩ Auto range conversion ±3%
Alarm range: 0-200MΩ any value
Testing time: 1-99s continuously adjustable
(3)Earth continuity test Earth continuity test
Measuring current: 10-25A/AC Constant current ±3%
Measuring range: 0-200 mΩ ( >10A ) ±3%
0-500 mΩ ( ≤10A ) ±3%
Automatic range switching
Alarm range: 0- 500 mΩ any value
Alarm range: 0- 500 mΩ any value
Alarm range: 1-99s continuously adjustable
(3) Ground continuity test
(3) 500 mΩOutput Voltage: ≤6V
Test Time: 1-99s continuously adjustable
Output Frequency: 50/60Hz two steps
Driving Method: Programmable Linear Amplifier The following (4) (5) (6) and the above side-by-side design
(4) Leakage Current test Current test
Current measurement range: 0-20mA with automatic range switching ±3%
Current alarm range: any value within 0-20mA
Phase switching: automatic
Voltage measurement range: 0-300V ±3%
Output voltage: 0-250V continuously adjustable
Testing time: 1-99s continuous
Testing time: 1-99s continuously adjustable
Output capacity: 1/2/3/5kVA Multiple specifications
(5) Start-up test Start-up test
Measuring current range: 0-20A ±1%
Current alarm upper and lower limits: 0-20A arbitrary value
Measuring voltage range: 0-300V ±1%
Measuring voltage range: 0-300V ±1%
Output voltage: 0-250V
Continuously adjustable test time: 1-99s continuously adjustable
(6) Power testPower test
Power measurement range: 0-6000W ±1%
Power alarm upper and lower limits: 0-6000W arbitrary value
The power of the test is measured by a voltage range of 0-300V. Voltage measurement range: 0-300V ±1%
Output voltage: 0-250V
Continuously adjustable test time: 1-99s
Continuously adjustable output capacity: 1/2/3/5kVA Multiple specifications