I, the main technical indicators
1, the maximum test force: 5000N
2, the accuracy of the test force measurement range: 200 ~ 5000N
3, the accuracy of the test force measurement: ± 1%
4, the test force staging: × 1, × 2, × 5
2, in a vibration-free environment, no corrosive media around and electromagnetic interference
3, the fluctuation range of the power supply voltage should not exceed ± 10% of the rated voltage
4, the equipment must be reliably grounded
Third, the operation and use
1, the system is powered on, preheating for about 15 minutes. Zero by load, press the calibration key, observe whether the calibration value for the measurement of the calibration value (annual inspection to determine the value), otherwise need to be adjusted, so that the display value and the measurement of the calibration value is consistent. Press the calibration key again.
2, the operation of the operating panel "up" or "down" and "stop" button to adjust the moving beam to the appropriate position. Note: After the operation of the "up" key, you must press the "stop" key before you can press the "down" key, and vice versa.
3, the specimen will be clamped on the auxiliary equipment, the upper end of the fixed firmly, press the load zero key, and then fixed the lower end, selected gear, and then selected speed.
4, press the "rise" key, that is, to start the test, tensile specimen fracture moving beam will automatically stop. Sometimes need to quickly press the "stop" key to move the beam to stop moving.
5, repeat the above 2, 3, 4 steps, continue the following test.
6, press the F3 key to print the test data.
Four, notes
1, the test process, should pay attention to the position of the moving beam, please ensure the best test space.
2, screw and inverted column should be lubricated every month, often check the lubrication condition.
3, please pay attention to keep the testing equipment clean, the site should be cleaned up after the test.
4, equipment operation during the abnormal phenomenon, immediately press the emergency stop (mushroom head) button.
5, reducer work 3800 hours should be replaced after the lubricant, lubricant grade G-N320.
I hope to be able to help you Oh!