1, in the hanging installation of shadowless lamp, because the product is suspended load-bearing equipment, so the ceiling or hanger bearing requirements, must be able to withstand a load of 500Kg.
2, no structural beams in the center of the operating room, and the roof for the concrete structure, concrete grade must be greater than 150, floor thickness greater than 10CM:
3, the installation process should be connected to the electrical wiring (power supply connected to the AC 220V), the link parts must be tightened. In the installation of shadowless lamp before the application of a 250V/10A circuit breaker to open the control circuit on and off.
4, LED shadowless lamp installation process must be strictly calibrated with the level of the hanger level.
5, ceiling height greater than 3.2m, should be added to connect the long frame.
The above is the answer provided by Mega Optoelectronics for you, I hope it will help you!