The warehouse temperature, humidity should be consistent with the operation of medical equipment instructions or labeling requirements.
Check the range of temperature and humidity settings in the warehouse, and confirm on-site the temperature and humidity of the warehouse and its storage of medical devices
Device instructions or labeling requirements are consistent.
Special temperature and humidity storage requirements for medical devices, should be equipped with effective control and monitoring of temperature and humidity equipment
or instrument.
Storage such as storage of medical devices with special requirements for temperature and humidity, focus on checking whether the storage room is equipped with and use
can effectively monitor and regulate the temperature and humidity of the instrument or equipment.
Wholesale of medical devices that require refrigeration, freezing storage and transportation, should be equipped with the following facilities and equipment:
(a) its scale of operation and business varieties of cold storage;
(b) used for cold storage temperature monitoring, display, recording, regulating, alarming equipment;
(c) to ensure that the refrigeration equipment for normal operation of the facility (such as standby) Generator set or double circuit power supply system);
(d) the need for transportation enterprises, should be equipped with refrigerated
cars, insulated trucks, or refrigerated containers, insulated boxes and other equipment according to the appropriate scale of transportation and transport environmental requirements;
(e) special low-temperature requirements for medical devices, should be equipped with facilities and equipment to meet its storage requirements.
On-site focus on checking the warehouse and its refrigerated, frozen storage, transportation facilities and equipment are equipped with the above
Full version : pharmaceutical cold storage installation process, medical device cold storage installation company, <<Medical Device Cold Chain Transportation & Storage Management Guide>>