▲ JHM-2028 main technical parameters
◆ Volume (L×W×H): 370mm×340mm×1570mm
◆ Weight: about 90kg
◆ Power supply voltage: AC220V±10%
-Frequency: 50HZ ~ 60 HZ
-Power: 1500W
-There are two types of pumps, one is the "Balance Pump" and the other is the "Hemodialysis Pump Head". 1500W
◆ Blood pump:
-Flow rate: 15 ~ 340 ml/min (tube I.D. Ф 6mm)
20 ~ 460 ml/min (tube I.D. Ф 8mm)
-Backup pump with the same data as the blood pump
◆ Heparin pump:
-Flow rate: 0.1 ml/h ~ 10 ml/h (±5%). h(±5%)
-Resolution: 0.1ml/h
-Syringe size: 20ml/30ml/50ml
◆ Arterial Pressure:
-Display range:-300mmHg~+300mmHg
(±10mmHg)
◆ Venous Pressure:
-Display range:-300mmHg~+300mmHg
(±10mmHg)
◆ Heparin Pump:
-Display range:-50mmHg~+300mmHg
(±10mmHg)
◆ Dialysate flow rate:
-300ml/min~800ml/min
Linearly adjustable (+10%-5%)
-Temperature:
35.0℃~39.0℃
-Resolution:0.1℃
-Conductivity:12.5mS/cm~15.5mS/cm
(±0.1 ms/cm)
◆ Ultrafiltration flow rate range: 0~1800ml/h(±30ml/h )
0~4000ml/h(optional)
◆ Dry Ultrafiltration flow rate range:0~2000ml/h(±30ml/h )
◆ Dry Ultrafiltration flow rate range:0~4000ml/h(optional)
◆ Dry Ultrafiltration flow rate range:0~2000ml/h(±30ml/h )
0~4000ml/h (±0.1ml) 2000ml/h(±30ml/h)
◆ Transmembrane pressure
-Range:-100mmHg~+600mmHg
(±20mmHg)
◆ Blood leakage monitoring:
-Blood leakage is more than 1ml per liter of dialysate (flow rate 500ml/min)
◆ Level monitoring: Using ultrasonic sensing transmission to monitor the level. Liquid level monitoring: ultrasonic sensing transmission
◆ Bubble monitoring:
-Infrared ray, reaction threshold:
At 200ml/min blood flow, there is a single volume of 200μl bubbles
◆ Inlet water pressure: 0MPa~0.6MPa
◆ Inlet water temperature: 5℃~30℃
◆ Ambient temperature: 10℃~30℃
◆ Inlet temperature: 5℃~30℃
◆ Inlet temperature: 5℃~30℃
◆ Inlet temperature: 10℃~30 Ambient temperature: 10℃~30℃,
- Relative humidity≤70%
◆ Cleaning/disinfection: chemical disinfection (citric acid, peroxyacetic acid, oxalic acid)
◆ Thermal disinfection: 80℃
◆ Backup power supply: it can be turned on automatically for 15-30 mins after the power failure. (Optional accessories)
▲ JHM-2028 Functions
Hemodialysis, hemoperfusion, plasma exchange, simple ultrafiltration hematochemistry treatment items; in the JH-2000 based on the new addition of the touch screen, with sodium ion curve and ultrafiltration curve function; in addition, you can also choose to configure: backup power supply, KT / V value, HCO3 adjustment, single needle dialysis, acetate The system can also be configured with backup power, KT/V value, HCO3 adjustment, single needle dialysis, acetate dialysis.
▲Clinical Indications of Each Function
◆ Indications of Hemodialysis: Same as JH-2000
◆ Indications of Simple Ultrafiltration: Same as JH-2000
◆ Indications of Hemoperfusion: Same as JH-2000
◆ Indications of Plasma Exchange: Same as JH-2000
◆ Sodium Curve and Ultrafiltration Curve Function:
◆ Sodium Curve and Ultrafiltration Curve Function:
◆ Sodium Curve and Ultrafiltration Curve Function:
- Applicable to reduce the incidence of hypotension, muscle spasm, headache, nausea, vomiting and hypertension during and after dialysis.
◆ KT/V Value Function:
- It can reflect the adequacy of dialysis of the patient by KT/V value. (Optional)
◆ HCO3 Regulation Function:
-Applicable to the patients with serious acidosis; it can correct the acidosis state better according to the patient's condition. (Optional)
◆ Backup Power:
-Automatically turn on for 15-30 minutes after power failure. (Optional)