How to control the shutdown of Siemens 16 row ct water cooler

1. Turn off the power of the compressor motor and stop the compressor.

2. Turn off the power of the chilled water pump to stop the chilled water pump.

3.) Turn off the power supply of the cooling water pump to stop it.

4, turn off the cooling water tower fan power supply, make the water tower fan stop running.

5. Turn off the power supply of each air-conditioning fan motor, and the whole machine stops working.