This equipment is divided into vacuum, warming, sterilization, exhaust, drying which stage
If in the vacuum warming stage, waiting for electricity, water when the exit to re-sterilize
If in the exhaust and drying stage. If you wait for electricity and water to exit, you do not need to re-sterilize. The sterilized items may be damp.
If in the sterilization stage
If it is a short power outage, water outage. In the sterilization does not affect the time again to get electricity, water, can continue to use to the end.
If it is a long time power outage, water outage. Maintain compressed air pressure (compressed air pressure to maintain the seal of the door tape, I remember
wrong words).
1 Watch the pressure change in the inner chamber, if it continues to rise, cut off the compressed air. Next time, change the door tape
2 Watch the pressure change in the chamber, if it continues to fall, wait for electricity, water, exit the program. Re-sterilize.
In a word, safety first. Pay attention to the pressure gauge pointer readings (chamber, jacket). Take appropriate measures.