The second level of biological laboratory must be equipped with equipment is

Second-level biological laboratory must be equipped with equipment is introduced as follows:

(1) laboratory table: whether it is the second level of biosafety laboratories, or other types of laboratories, the laboratory table is indispensable, the relevant equipment and supplies are placed here. Its countertop material to acid and alkali resistance of the physical and chemical solid core board is appropriate, under the countertop drawers and cabinets to place the experimental supplies.

(2) safety cabinets: the safety cabinets for biological safety cabinets, according to national norms, the second level of biological safety laboratories must be configured with a safety cabinet. This is a kind of isolation equipment, the role is to prevent the proliferation of harmful suspended particles in the laboratory, can provide comprehensive protection of staff, experimental samples, the environment in the laboratory.

(3) interlocking transfer window: according to national norms, the second level of biosafety laboratories must be configured with two interlocking transfer window to ensure the safety of logistics. There should be ultraviolet lamps inside the transfer windows in order to sterilize the items entering the laboratory.

(4) interlocking doors: according to national norms, the secondary biological safety laboratory must be configured with a set of interlocking doors, interlocking doors, 1 door is not closed, it is not possible to open the other door, thus playing a buffer effect of airflow.

(5) automatic door closers: automatic door closers must be installed on the door of the secondary biological safety laboratory.

(6) Scrubbing table: in the secondary biosafety laboratory scrub disinfection room to be equipped with a scrubbing table, and the water should be used in the laboratory triple water, sinks and downstream hoses to be used in acid and alkali resistant PP materials.

(7) multifunctional microcomputer controller: the temperature and humidity display and control of the secondary biological safety laboratory, send and exhaust fan start and stop and start and stop sequence, lighting switches, ultraviolet lamp control, these are controlled by multifunctional microcomputer controller.

(8) micro-pressure differential meter: according to national norms, two biological safety laboratories must be equipped with a set of indoor and outdoor pressure differential amount of indoor and outdoor pressure difference, and control of differential pressure micro-pressure differential meter.

(9) ultraviolet lamp: biosafety laboratories must be equipped with 2 ultraviolet lamps in order to provide physical disinfection for staff.

(10) socket: the laboratory must have a socket, the second level of biosafety laboratories are no exception, to configure a sufficient number of nine-hole sockets and three-phase four-hole sockets to be able to do, and sockets are embedded design.