What's wrong with the ge logiq p5 abdominal ultrasound that cuts out when I use it? The probe sometimes doesn't recognize it either!8C, 11L, E8CS, 3SP

Have you used a voltage regulator? Look at the normal voltage is not normal, if you do not use a voltage regulator also measure the voltage of the wall socket is how much, if it is not a problem in the hospital, that is, the power supply of the machine has a problem, often power outages for no apparent reason is very easy to damage the machine's board, if it is the insertion and removal of probes do not recognize that there may be a bad contact, if it is said to be normal use such as the use of the high-frequency for abdominal, and then change back to the high frequency found that the non-recognition of the board is surely a problem. Board has a problem. Machine problems can only find manufacturers after-sales, if there is insurance that is not money, if you do not buy insurance that suggests that you take advantage of this repair first and they discuss the insurance to buy, and then the maintenance fee will be exempted, if it is really bad a power supply plus a board, then the maintenance fee of several tens of thousands of dollars at least

.