In the automatic pacemaker test method, seven parameters are measured semi-automatically: pulse frequency, beat width, beat amplitude, weak beat interval, strong beat interval, sensitivity, and A-V delay. Pulse width was measured in combination with PPM measurements. For each parameter, 10 different test values are preprogrammed. The user can selectively test several of the 10 values, test one or more values of the user's choice, or delete any of the tests.

User-modified values can be stored and used later as long as the DELTA 3000 is not shut down. The user may also not store modified values; thus, values that were not part of the original test will still be called up in future tests. The test program continues until all parameters have been tried, and the DELTA 3000 stores the measured values for each test in memory. At the end of the automatic test program, the test results are transferred to a printer or to a PC via the RS232 interface. The set values are compared with the measured values of the DELTA 3000-user-manual-R2.doc34 . If the deviation is more than 10%, a comment on the effect is issued and printed out.