The minimum effective usable area in the CT machine room is 30 m2, and the minimum unilateral length in the machine room is 4.5m. According to the above national standards, your driver's room is 6.5m * 5.8 m2=37m2, and the area and width of the machine room meet the national standards.
Extended data:
X-ray machine room construction cooperation project:
(1), peripheral requirements: the machine room is made of 20 hollow brick walls, embedded with water and electricity, leveled with mortar on the brick surface, with a layer of barbed wire, anti-radiation coating and white surface;
(2) Requirements for suspended ceiling: remove mortar from suspended ceiling, reinforce barbed wire with flat iron expansion screws, apply protective coating, and paint the surface with white ash or suspended ceiling.
(3) Ground requirements: The cable trench should be dug and the surface should be paved with bricks or other materials.
(4) Requirements for reserved holes in the operating room: reserved small doorways (width 0.9m * height 2. 1m) and observation windows (window sill height 0.85m, hole width 1.7m, height 1. 1m).
(5) reserved entrance hole of computer room: width 1.5m * height 2.2m ..
(6) Bury the upper-middle 15cm inside the gate into the distribution box (connected to the 1.5*2 line) as the working light.
(7), the door to the distribution box wiring (1.5*2 line) as an interlock device, the door has an interlock switch.
(8), responsible for metope, ceiling, floor decoration.