Autoclave sterilizer use:
1 Add the appropriate amount of water to the outer pot, put the items to be sterilized into the inner pot, cover the lid and symmetrically twist the screw.
2 Heating to produce steam in the pot, when the pressure gauge pointer reaches 33.78kPa, open the exhaust valve, the cold air out, at this time the pressure gauge pointer drops, when the pointer falls to zero, the exhaust valve will be closed.
3 continue to heat, the pot steam increase, the pressure gauge pointer and rise, when the pot pressure increased to the required pressure, the fire will be reduced, according to the characteristics of the sterilized items, so that the steam pressure to maintain the required pressure for a certain period of time, and then will be cut off or cut off the fire sterilizer, and let it cool naturally, and then slowly open the exhaust valve to exclude the residual gas, and then you can open the lid to take things.
Autoclave sterilizer precautions
1 items to be sterilized should not be placed too tight;
2 must be fully excluded from the cold air, otherwise the temperature inside the pot can not reach the specified temperature, affecting the effect of sterilization;
3 sterilization is completed, can not be deflated to reduce the pressure, otherwise the liquid in the bottle will be violently boiled, washed out of the bottle stopper and overflowing or even lead to the container burst. To wait until the pressure inside the sterilizer to equal the atmospheric pressure before you can open the lid;
4 microcomputer automatic control of the autoclave, just let go of the cold air, the instrument can be automatic constant pressure timing, time to automatically cut off the power supply and siren, it is very convenient to use.