My wife works on the second floor, and the radiology department’s photography room is downstairs. Is there any problem?

1. When designing and constructing the computer room of the radiology department, protection issues must first be considered, and protective measures have already been taken. According to relevant national standards, the computer room of the radiology department has strict requirements for the area, floor height, and wall protection capabilities of the radiography room for X-ray machines of different capacities. For example, the wall should have 2mm lead equivalent in the main ray direction. Protection thickness, other walls should have a lead equivalent protection thickness of 1mm.

2. After installation, the equipment in the radiology department needs to pass a protection assessment before it is officially put into use. The testing institutions include local metrology institutes, prefecture-level health law enforcement detachments, etc., only Only after passing the inspection can it be put into use.

3. Radiological medical equipment and protection need to be tested at least once a year.

Based on all the above aspects, there is no need to worry about ionizing radiation whether in the room above the photo studio or next door.