Computerized intermediate frequency therapy instrument is useful

Computerized intermediate frequency therapeutic instrument for therapeutic rehabilitation physiotherapy health care of an advanced medical equipment, this machine adopts the microcomputer control technology, composed of medium and large-scale integrated circuits, equipped with 8 can be synchronized and asynchronous output channel, 4-way thermoelectric output channel, 2-way interference with the electric channel, 2-way ion introduction channel.

1. Insert the net power connector of the power cord into the power input socket of the equipment on the electrotherapy machine, and the plug into the fixed net power socket.

2. Place the electrode plate in the affected part of the patient, between the electrode plate and the human body should be dry towel as a mat, the thickness of which depends on the treatment site, generally 1-2 cm. the more back the mat, the better the thermal effect of the deep tissue. The affected part should be placed between the two electrodes as far as possible when using the device. In order to prevent electrode displacement, the electrode and pad can be tied to the affected part with a dry bandage.

3. Select the "treatment time" (timer time) according to the treatment needs.

4. Turn the "Output Selection" to "Preheat".

5. Press the "power switch" from "off" to "on", the "switch" green light, the machine power on, the electronic tube is connected. The power supply in the machine is connected, the electronic tube is lit, start "preheating".

6. Turn the "power supply voltage adjustment" to the voltmeter indicator for 220V. "Preheat" for two minutes.

7. Rotate the "Output Selection" to "Treatment 1" file, then the yellow light is on, the machine began to work normally. The ammeter indicates the size of the oscillating tube screen current.

8. Turn the "output tuning", the ammeter indication also changes with the resonance point, when the resonance point is reached, the ammeter indication of the maximum. Treatment is best done at the resonance point.

9. According to the needs of the treatment, you can choose different values of current. Such as "output selection" of the "1" file is not enough dose, you can choose "2", "3", "4" file. "4" gear to increase the dose. Each time you change gears, you should re-tune to make it reach resonance.

10. When the treatment time is up, the output is automatically cut off, the ammeter indicates zero, the "output selection" should be returned to the "preheat" in order to turn off the power supply, and then remove the electrode plate from the affected part of the patient.

11. If the machine needs to be used continuously, when replacing the patient, just rotate the "Output Selection" to "Preheat", the yellow light will not be on and the ammeter has no indication, then you can replace the patient, place the electrodes, and then repeat the above steps.

12. This machine is equipped with a timer, with "normally open" and "timed" function. Do not need to "timer", you can "treatment time" in the "normally open". If you need to control the "treatment time" automatically, you can place the "treatment time" at the desired position each time you use it, and then the output can be cut off automatically, and the machine will return to the "warm-up" state. After replacing the patient, only need to rotate the "treatment time", the machine has to restore the output.