Color ultrasound, biochemistry, blood cell analyzer, plasma collector, electrolyte analyzer, chemiluminescence immunoanalyzer, CT and nuclear magnetic **** vibration of the workstation and other such instruments basically single-phase 3,000VA UPS UPS is enough, part of the models need to be used for large power to a single-phase 6,000VA is also enough. Single-phase price is cheaper.
CT and nuclear magnetic **** vibration of the machine need to use three-phase four-wire high-power frequency machine uninterruptible power supply, the power is basically in the 80KVA ~ 160KVA, of which the CT machine with 120KVA more, nuclear magnetic **** vibration with 100KVA and 160KVA more.