. About the relevant unit of ozone therapy (1) At present, the domestic ozone concentration is recommended to use ug/ml to express. (2) Volume unit of gas produced by ozone generator: ml. (3) Time unit of ozone generator's timed gas taking function: min. 2. Output concentration of ozone (1) The maximum output concentration of ozone is 60ug/ml. (2) The ozone concentration adjustment is continuously adjustable, and the concentration order
1. Ozone concentration is recommended to use ug/ml to express.
(2) The volume unit of gas produced by ozone generator: ml.
(3) The time unit of ozone generator's timed gas taking function: min.
2. Output concentration of ozone
(1) The maximum output concentration of ozone is 60ug/ml.
(2) The adjustment of ozone concentration is adjusted continuously, and the difference of the concentration order is 1ug/ml.
(3) The error between the apparent output concentration of ozone and the actual output concentration is ≤±8%.
3. Ozone generator line output pressure,
Output pressure >1bar (1bar=10Pa) to prevent the occurrence of backsiphonage.
4. Ozone absorption device
Ozone generator should have ozone absorption function for absorption and degradation of excess ozone.
5. Pipeline pressure detection and alarm system
Pipeline pressure detection and alarm system is used to detect the pressure of the inlet pipeline, the internal pipeline, the output pipeline, and is used to detect and warn the airway pressure is too large or too small, as well as disorders, to prevent overpressure and backsiphonage.
6. Program-controlled ozone generator self-test function
Self-test function is used to detect the start-up state of the equipment and abnormalities in the work, to facilitate the medical staff to carry out normal medical work, to prevent medical accidents, and to facilitate the maintenance of remote equipment maintenance personnel.
7. Ozone generator gas source
Must be a high-pressure container bottled medical oxygen, medical oxygen standards should be in line with national standards for medical oxygen.
8. Output gas impurity requirements
The main component of the output gas is ozone and nitrogen-containing compounds such as nitrous oxide concentration should be
9. Medical ozone generator medical range of regulations apply
Medical ozone generator must be strictly limited to clinical use in the State Ministry of Health and Food and Drug Administration within the limits. Specifically for the three types of ozone generators limited scope of application must be used within the specified range, not limited to the scope of the ozone generator can not be carried out by the State Ministry of Health and Drug Administration to allow all therapeutic projects.