Siemens g20 ultrasound can not turn on

Siemens g20 ultrasound can not boot into the repair mode, press the menu key, the password is empty, self-test once to see if there is no obvious error message. Then go into EVENTLOG, plug in a USB flash drive and hold down CTRL + ALT + S to download the error log, analyze the internal TXT file, there will be a detailed description of the reasons for the error, can be used as a basis for reference, the long false ni is the HV error or TI error; first suspected of HV problems, high voltage is supplied with the DCDC board, the left side of the machine is the vertical board of the DCDC board, 48V Input, HV is 100V, with a meter to measure the dynamics of 100V, when the machine appeared after the error 100V are always there.

Siemens color ultrasound is a high-definition black and white B ultrasound plus color Doppler, first of all, ultrasound frequency shift diagnostic method, that is, D ultrasound, this method applies the principle of the Doppler effect, when the source of the sound and the receiver (i.e., probe and reflector) when there is a relative motion between the frequency of the echo has changed, the frequency of this change is called the shift, D ultrasound, including pulsed Doppler, continuous Doppler, and color Doppler blood flow Images. The direction of development of ultrasound in the medical field is color ultrasound, commonly used in obstetrics and gynecology, prostate and seminal vesicles, small organs, abdominal organs and vascular disease.