(A) the use of
1, to confirm the correct connection of the power supply, the power switch on the right side of the sterilizer will be turned on, and then press the "POWERON/OFF" button on.
2, make sure the pressure inside the pot is 0, gently press the handle of the sterilizer lid downward, and open the lid by pressing the lid lock release pedal. Fill the sterilizer pot with water (distilled water) until you see the water spread over the horizontal bar in the center hole of the sump (approximately 2.0 liters of water is required).
3. Place the items to be sterilized, close the lid to the magnetic seal and lock the lid in place. Be careful not to overfill the lid with individual small pellets.
4, select the operating mode (generally are selected ordinary sterilization mode, select other modes need to ask the safety officer).
5, start the sterilization program:
(1) to determine the water level in the bleeder bottle between "HIGH" and "LOW".
(2) Make sure that the water level in the drain bottle is low enough that it will not touch the top of the vent pipe.
(3) Press "START/STOP" to start the sterilization process.
6, sterilization is complete, the temperature drops to below 90 degrees (pressure inside the pot is 0) before you can open the lid and take out the sterilized items
7, press the "POWERON/OFF" button to disconnect the power supply.
(2) Precautions
1, before the pressure reaches 0MPa, do not open the lid of the sterilizer.
2, the pressure gauge should be stopped when there is an abnormality.
3, open the lid of the sterilizer pot, should pay full attention to the steam from the sterilization chamber to prevent scalding.
4, when placing the sterilized items should be careful not to touch the temperature probe in the damage liner.
5, when the water level in the bleeder bottle is higher than the "HIGH" mark, the water will be poured out to the "LOW" mark before sterilization.
6, when the water level in the pot is lower than the crossbar in the center of the chassis, you should add appropriate amount of distilled water to the crossbar can be diffused.
7, in order to avoid blocking the pipe system, should be changed frequently; the equipment is ready for a long time out of service, the sterilization pot to empty the water.
Related description of the autoclave
Autoclave is also known as high-pressure steam sterilizer, can be divided into portable sterilizer and vertical autoclave. The use of electric wire heating water to produce steam, and can maintain a certain pressure of the device. Mainly has a can be sealed barrel, pressure gauge, exhaust valve, safety valve, electric heating wire and other components.
Safety valve: over the rated pressure, release pressure.
Hand wheel: rotating compass type to open the lid door, simple and convenient (suitable for products over 40 liters).
Pressure gauge: Indicating pressure display.
Sealing ring: self-rising sealing ring; bottom movable wheel.
Vapor collection bottle: 3L, made of polyethylene; with product medical device registration certificate, each through the Technical Supervision Bureau strict testing (each device with a set of test certificate).