AGFA equipment failure analysis

Failure analysis and solutions

Failure 1: Repeat Printing

Phenomenon: the same patient with a piece of film more than one or several copies (in the case of the print setup is not wrong), the camera screen without error prompts, the piece out of the control panel (Impax MG3000) on the display of the number of images does not reduce.

Possible causes and solutions: This is because the camera host controller part does not receive and should have received the instruction to erase the information, the reason for the sewer clogging is most likely. Check to see if the film washer sewer pipe is clogged with dirt or the sewer pipe is deformed or bent under pressure, and resolve as appropriate.

And redundant images can be deleted from the control panel: the method is to first press F1 to display the unprinted image code in the host computer, select the image code to be deleted through the up and down Ken, press the ERASE key, and then press the STORE B &W Ken to confirm according to the prompts.

Failure 2: Continuous blank sheet

Phenomenon: Continuous printing of one or more blank sheets without images, the camera screen no error message.

Possible causes and solutions: 1, the supply box film due to moisture sticking to each other, the camera suction cup each time to suck more than one film. The solution depends mainly on the camera to maintain the working environment, air conditioning, dehumidifier is essential. Indoor 24 hours must maintain a constant temperature of 20-22 degrees, humidity control at about 40%, stock film indoor preservation environment is the same. 2, driven by the suction cup work belt is too loose or too tight. Open the front cover of the laser camera, slightly pull out the camera printing system, check the tightness of the belt, adjust the elastic dowel to the belt tightness.

Fault three: double print

Phenomenon: each time the second print the same image, the LCD display LEVEL DRIP PAN, suggesting that the liquid level in the developing barrel is too low.

Possible causes and solutions: 1, immediately check whether the connection of the plastic hose is too bent, as far as possible so that the hose is short and straight. 2, check the supply of developer barrel developer is too little, add developer.

Failure 4: When each time after cleaning the film washer, in the development tank and fixing tank added to the scale above the development, fixing liquid, after power on, the camera can not work, the night crystal display DEVP LEVEL or FIXER LEVEL, suggesting that the development or fixing liquid surface in the film washer is low.

Possible causes and solutions: 1, check whether the hose connecting the developer and fixer tanks with the developer and fixer drums is loose, or whether the fastening parts are loose. In general, this will not be a problem. 2, open the back of the machine on the cover, check whether the display, the liquid level next to the fixing tank detector head is covered with dirt, remove the detector head and gently wipe clean with a soft cloth, re-installed can be. 3, the above two methods can not be resolved, you can manually developer or fixing liquid into its slot to the fast full slot, you can cope with the work immediately, but can not be a fundamental solution to the problem.

Failure 5: insurance jumped off the beginning

Phenomenon: This is the case every time after the change of the drug water and show that the fixing back suction bad situation, the LCD display SERVICE 502 or 504, prompting the developer (502), fixing (504) temperature is too low in the film washer, film washing machine can not work properly. If the camera is in normal operation, suddenly appears LCD display SERVICE 501 or 503, then prompts the development (501), fixing (503) temperature is too high, the film washer can not work properly. <br>

Possible causes and solutions: according to the fault four processing, open the front of the machine cover, find the dark red (developing), blue (fixing) insurance switch button, use a screwdriver to press it down, restart the camera can be. Developing or fixing temperature is too high, you can artificially release 20-30% of the developing or fixing fluid, let the machine and then automatically replenish some of the low-temperature developing or fixing fluid.

Failure 6: low liquid level in the washing tank

Phenomenon: the camera does not work after power on, the LCD display WATER LEVEL, suggesting that the liquid level in the washer is too low.

Possible causes and solutions: 1, it is likely to forget to open the water supply faucet switch. 2, low tap water pressure, check and clean the tap water filter, unclogging the connection to the water supply faucet hose. 3, if the time is very urgent, but also by switching on the camera once more to deal with the work.

Fault VII: Camera transmits part of the card

Phenomenon: Often you can hear the abnormal ringing of the card, LCD display SERVICE 654 or SERVICE 850, prompting the camera to transmit part of the card.

Possible causes and solutions: 1, SERVICE 654 prompted by the camera suction card parts of the card, open the camera front cover, slightly pull out the camera printing system, from the bottom of the card can be extracted. 2, SERVICE 850 prompted by the camera back to transmit part of the card, open the camera back cover, find the card card can be managed to pull out. 2, SERVICE 850 prompted by the camera back to transmit part of the card, open the camera back cover, find the card card can be pulled out.

Failure 8: card washing machine parts

Phenomenon: carefully hear the washing machine parts of the abnormal sound, the screen LCD display SERVICE 506, prompted to drive the washing machine motor speed is not normal. <br>

Possible causes and solutions: common for the developer, fixing, washing, drying parts of the card, open the machine cover, find the card film and try to pull out. It is also possible that the 锟shaft of the film washer is not round or bite gear damage, it must be replaced.

Failure 9: camera LCD panel does not show

Phenomenon: the camera screen LCD panel does not light, and the film washer Kun shaft rotation sound.

Possible causes and solutions: check whether the front cover of the camera is tightly covered, after the cover is tightly removed immediately.

The above faults as long as the imaging technicians in the work of careful observation, humbly ask for advice, can completely grasp the way to exclude, to bring great convenience to our work. If you do not understand, you have to contact the manufacturer, such as manufacturers of engineers to come, may be two or three days of work can not be carried out, to the daily work has a great impact. Of course, now the structure of large instruments is very complex, some rare faults or electrical components wai off, our ordinary technicians still can not do anything about it, we have to immediately look for professional engineers to repair.