I. Requirements for laboratory space
1. The width of the main entrance of the laboratory shall not be less than 80cm, and the height of the door shall not be less than 198cm.
2. The computer room has an area of at least 3×3=9 square meters. Leave at least 0.5m space around the host for future maintenance.
Gallios/Navios mainframe (width x depth x height = 102 cm x 70cm x 6 1 cm, weight = 109 kg)
Gallios/Navios air pump box (width x depth x height = 7 1 cm x 58cm x 46cm, weight = 59kg)
Second, the laboratory power requirements
1. The laboratory needs to purchase 3KVA online uninterruptible power supply, and Sander UPS is recommended.
2. The installation of the instrument needs to use at least two 220V, 16A, 1, 10A sockets, which are respectively installed on the wall behind the instrument.
3. The instrument must be well grounded, with zero grounding voltage lower than 1VAC and grounding impedance less than 4 ohms.
4. Equipped with two high-quality commercial multifunctional wiring boards for supporting equipment.
Third, the laboratory temperature requirements
15.5 ~ 32℃, the temperature change shall not exceed 2℃, and the air conditioner shall not use the same power cord as the instrument.
Four, laboratory humidity requirements
20% to 80%, no condensation.
V. Laboratory facilities
1. Two stable desks are needed, one for the instrument host and the other for the computer, monitor and printer.
Table size:1.5m (width) × 0.8m (depth) × 0.8m (height)
2. Prepare at least 1L distilled water or "Milli Q" purified water.