Parameters and sources of AGSS waste gas treatment station

http://6 1 . 49 . 18.65/public files/business/CMS resources/moghcws/cmsrsdocument/doc 6905 . doc

The manufacturing standard of natural gas terminal shall be BS EN 737- 1.

The manufacture of gas quick plug follows BS 5682 standard.

The processing and manufacturing standards of AGSS shall conform to BS 6834 standard.

The application of gas terminal conforms to HTM2022 standard.

It can provide seven kinds of medical gas terminals and quick plugs, such as oxygen, negative pressure, compressed air, nitrous oxide, nitrogen, carbon dioxide and AGSS. Different gases have different sockets.

There are two kinds of gas terminal connectors, one is φ 10 copper pipe and the other is hose connector.

Use ISO32 color standard for gas identification.

Provide mounting base and plastic ferrule to help assembly.

Provide matching wall-mounted plastic boxes, which can be installed openly or concealed.

The rated pressure of the place where medical gas is used.

(kPa) Typical use process

(L/min) Design flow

(L/min) Test flow

(liter/minute)

Medical air operating room 400 20 40 80

Keywords intensive care unit, newborn,

Intensive care unit 400 60 80 80

Other ward beds 400 10 20 80

Instrument air and nitrogen orthopedics and neurosurgery operating room 700 350 350 350

Medical vacuum major surgery 40(300mmHg

Vacuum pressure) 15-80 80 80

Minor operation, 40 beds in all wards (300mm Hg

Vacuum pressure) 15-40 40 40

N2O4005-10100100 was used in oxygen operating room and anesthesia point.

All other wards use point 400 5 10 40.

Nitrous oxide surgery, obstetrics, all wards use 400 points.

5- 10 15 40

Nitrous oxide/oxygen

Mixed gas LDRP (delivery, delivery, recovery, postpartum)

Room number 310-400 5-10 275 275

Dental and all other wards beds 400 5- 10 20 40

Helium/oxygen mixture intensive care unit 400 40 100 100

Scores of carbon dioxide operating room, contrast room and abdominal examination 400 5 20 80

Exhaust gas from anesthesia or breathing is discharged to operating room, anesthesia room, ICU15 (113mmhg.

Vacuum pressure) 50-80 50-80 50-80

Department unit oxygen vacuum medical air nitrous oxide/oxygen mixture nitrous oxide anesthesia or breathing waste gas nitrogen/instrument air carbon dioxide helium/oxygen mixture

Endoscope/cystoscope in operating department131-111/a-

Main operating room 2 32-2111a-

Auxiliary operating room 221-11-1a-

Orthopedics/neurology operating room 241-1121a-

Anesthesia room111-1-

Recovery room 2 2 1-

Outpatient operating room 2 2 1-

Delivery room1111-

Delivery room 2 2 1 1-

Postpartum recovery 1 2 1 1-

Baby room 1 1 1-

Pediatric neonatal intensive care 222-

Pediatric intensive care 222-

Nursery 1 1 1-

Pediatric ward 1 1-

Diagnostic EEG, ECG, EMG 1 1-

Digital subtraction angiography room (DSA) 222- 1A 1A6-

MRI 1 1 1------

Cat room 1 1 1-

Ophthalmology, Otolaryngology, ENT-11-

Ultrasonic 1 1-

Endoscopy 1 1 1-

Urography 1 1-

Linear accelerator 1 1 1-

Ward and other wards 1 1A 1A-

Psychiatric ward-

Burn ward 2221a1a1a65438-

Intensive care unit 222-1a-1a

CCU 222-- 1a---

Emergency room 2 2 2-

Dialysis 1 1 1-

Trauma treatment room 1 2 1-

Examination/treatment/disposal 1 1-

Gypsum house111a-1a-

Animal research121-1a1a1a65438-

Autopsy 1 1- 1A-

Cardiac catheterization 2 2 2-

Disinfection room 1 1×-

Dentistry and Oral Surgery11-1a-1-

General outpatient service 1 1