Ultrasonic height and weight scale height is not allowed how to debug?

With the arrival of the era of intelligence, ultrasonic height and weight scales have also come close to people, whether it is a hospital medical examination, or corner drugstore, or large shopping malls, there is an ultrasonic height and weight scales, which can allow people to grasp their height and weight information at any time.

Ultrasonic measurement of height and weight in general will not be a big deviation, when the deviation is large, not allowed when how to debug? Xiaoho on the above Wo technology ultrasonic height and weight scale as an example, to explain how to debug.

Zhengzhou on the Harvest Electronic Technology's height and weight scale in the factory are after layers of debugging, measurement data will not appear deviation, if there is a deviation, there are probably the following reasons:

1, the upper left corner of the circuit board position connected to the measurement of the probe's plug position, the plug whether there is a loose or poor contact;

2, the ultrasonic measurement height Whether the probe is pressed against the line;

3, boot the probe standing under someone;

4, height measurement probe is not tilted, whether there are other debris around the probe blocking;

5, the transportation process is damaged, resulting in inaccurate measurement results, or even can not be used.

How to solve?

1, use the key to open the back door, turn off the machine after the ultrasonic plug, re-plugging.

2, check whether the probe pressure line, if any pressure line to adjust.

3, reboot, boot time on the scale do not stand people or have anything.

4, in the transportation process of the damage, timely photo, contact the manufacturer, for replacement.