Canon detector specification parameters
Product specification CXDI-50G
Purpose: General radiological imaging.
Methods: Flat panel detector: scintillator and amorphous silicon.
Sensor: LANMIT (large area new MIS sensor and TFT)
Scintillator: GOS(Gd2O2S:Tb)
Pixel spacing: 160× 160 micron
Pixel: 2208×2688 pixels (5.9 million pixels)
Image size: the automatic measurement size can reach 35×1075px (14×17 in).
A/D converter: 14 bits
Gray scale: 4069 gray scale (12 bits)
Preview image access time: about 3-5 seconds after X-ray exposure.
DICOM: compatible with DICOM 3.0
Print management service class (SCU), storage service class (SCU), etc.
Voltage: AC 100-240 volts, 50/60 Hz.
Energy consumption: maximum. 170VA
Working environment: sensor unit: 5-35℃(4 1-95℉), 30-50% relative humidity (no condensation).
Certification: FDA 5 10(k). Fcc class a. UL 260 1- 1。 6060 1,CE0 197。
Size: sensor unit (width× length× length): 480× 48/kloc-0 /×15mm (18.9× 18.9× 0.6 inch. )
Weight: sensor unit (excluding cable): 3.4 kg (7.5 lb)
Standard accessories: sensor unit, power box, remote control switch, X-ray interface cable.