What principles should be followed in the design of water purification equipment

Good design can guarantee the stable operation of the equipment, to provide pharmaceutical companies with high-quality standard production water, to meet the latest version of the GMP standard requirements, in addition, the water purification equipment is also based on the actual needs of the user design.

Water purification equipment designed to follow the principles

1, the whole set of devices to follow a program operation. Even if there is no water demand, the purified water preparation system also maintains the cycle of operation.

2, RO device with self-circulating pipeline, in the tank level is full, do not need to enter the water, the device automatically switches to a small cycle of operation, the second RO water back to the intermediate water storage tank.

3, the conductivity of the first and second RO water, the PH value of the first water for continuous monitoring, with unqualified water discharge and automatic alarm function.

4, the system for 24h uninterrupted operation, no water demand to maintain low-flow circulation in the water pipeline to prevent dead water.

5, in order to facilitate the verification of the quality of water from each treatment unit, to confirm the performance of the single unit of equipment, each unit of equipment set up sampling points.

6, online monitoring of RO water temperature, flow, pressure and water conductivity, pH, flow and pressure, RO water temperature, water conductivity, pH, etc., with a record, not less than 6 months storage space and print function.

7, system disinfection: the system uses pasteurization; configuration of independent cleaning system, pure water distribution system design pasteurization.

8, the safety of the equipment is reliable, in the first place to protect the operator under the premise of protecting the safety of the equipment, the safety hazards at the installation of protective devices and warning signs.

9, all contact with product water parts using 316L stainless steel, pure water tank mechanically polished inner surface Ra<0.4μm, matte finish on the outer surface.

In addition to the overall configuration and structural design of the equipment to follow the relevant principles, the design and installation of the water quality delivery pipeline should also avoid dead ends, blind tubes, interfaces and elbows at the smooth and flat, pharmaceutical water purification equipment out of the water quality is not subject to secondary contamination, the safe supply!