30cm×30cm×25cm
The volume of the object package with the under-vented pressure steam sterilizer shall not exceed 30cm×30cm×25cm. the number of layers of cloth package shall not be less than two, and the volume of the object package used for the pre-vacuum and pulsating vacuum pressure steam sterilizer shall not exceed 30cm×30cm×50cm.
Additionally the use of The weight of the metal packages shall not exceed 7kg, and the dressing packages shall not exceed 5kg. it should be noted that the loading capacity of the lower exhaust sterilizer shall not exceed 80% of the capacity inside the cabinet? ; Pre-vacuum sterilizer loading should not exceed 90% of the cabinet chamber volume.
Expanded Information:
Pressure steam sterilizer: according to the discharge of cold air in different ways and to different degrees, is divided into the lower exhaust pressure steam sterilizer and pre-vacuum pressure steam sterilizer two categories.
Lower exhaust pressure steam sterilizer is the use of gravity replacement principle, so that the hot steam in the sterilizer from the top down, the cold air from the lower vent discharge, the cold air discharged by the saturated steam instead of the use of steam to release latent heat to make the items to achieve sterilization.
In addition to this, it should be noted that the sterilized items can be disassembled must be disassembled, such as syringes for packaging, the core should be withdrawn. Exposure of all surfaces of the item, such as scissors and vascular clamps must be fully propped up) in order to facilitate the sterilization factor to contact all surfaces of the object.
Shaoxing central blood station - sterilization technical specifications