Image blue no image, cable damage, replace the cable
Image reddish, the CCD is damaged, replace the CCD
Image reddish, the color corrective color glass is damaged, replace the color glass
Image interference, cable sheath damage and shielding cable damage, replace the cable
Focus is not clear, focusing mirror is damaged, focusing mirror is damaged >Image interference, cable sheath damage and shielding cable damage, replace the cable
Focus is not clear, focusing mirror damage, focusing mirror damage
Image reflections, horseshoe plate damage, repair horseshoe plate
Camera system specific problems and prevention are as follows:
I. Camera system common problem is the camera failure, failure phenomenon is often interference ripples, color streaks and other anomalies. It is recommended not to plug and unplug the camera from the host often, when not in use can be connected to the host, but at least once a week the surface of the work (can be wiped with a dry cotton swabs or erasers) camera cable and fiber optic avoid a large angle of bending, should be a natural arc state. Shake the camera line monitoring in the event of miscellaneous flickering or color change. Typical camera linkage damage. It is appropriate to replace as soon as possible to avoid burning the CCD chip.
II. Optical lighting system common problem of light-guiding beam skin breakage, thus accelerating the internal fiber breakage, affecting the light-guiding performance, but also due to skin breakage and leakage of light caused by local heating problems. This problem mainly occurs in laparoscopic surgery with long operation time, where the light guide beam is bent under force for a long time. However, by correctly holding the mirror, the problem of skin breakage of the light guide beam can be effectively controlled and the service life of the light guide beam can be prolonged. When using a 0° laparoscope, hold the mirror with the beam connector facing downward, and when using a 30° laparoscope, hold the beam in your hand in an arc from the mirror connector.
Common faults of camera system:
Camera host: blurred image, image color deviation, image flashing screen, no image output, the host can not be started.
Camera: image interference, image off-color, focus or zoom failure, loose bayonet, sheath fracture, cable damage.
Cold light source: no light source output, the host can not start, light source flicker, bulb does not light, timing alarm, etc..
Gas belly machine: fault code report error, unstable pressure, insufficient air pressure , no gas output, can not start, etc..
Light guide beam: cable skin damage, light guide fiber broken wire brightness is not enough, guide head burn damage, adapter damage, etc..
Camera system maintenance scope:
Camera host maintenance: maintenance of the image processing board, maintenance of the driver board, the host power supply maintenance, back-end output board maintenance.
Camera repair: camera cable repair, replace the camera cable, replace the camera CCD, repair the camera tuning mirror.
Cold light source repair: motherboard repair, high-voltage board repair, replacement of high-voltage module, replacement of light bulbs, light bulb timing module zero.
Air belly machine repair: repair control circuit, repair gas circuit, repair power board, repair and replace the control valve.
Guide beam repair: replacement of guide fiber, guide head, adapter, outer tube, etc..
Hangzhou Si-han Medical Equipment Co., Ltd. is a professional endoscopy system after-sales service provider, providing many such as Storz (STORZ), Stryker (Stryker), Wolf (WOLF), Schleifel (Smith& Nephew), Olympus (Olympus), Medtronic (Medtronic), Johnson & Johnson (Harmonic), Snake (Harmonic), and so on. Harmonic (Harmonic), Snake (Aesculap), CONMED (CONMED), Linvatec (Linvatec), RUDOLF (RUDOLF), XION (XiON), PENTAX (PENTAX), Arthrex (Arthrex), Valleylab (Valleylab), ERBE (ERBE), Bovie (Bovie), AMO, Alcon, Barschlomb, BienAir, ScanMed and other brands of equipment maintenance; focusing on the operating room camera system, power system handle, scleroscope, ultrasonic knife, ultramammography handle, high-frequency electrosurgical knife, pneumoperitoneum machine, cold light source, the mainframe of the camera system and other maintenance projects. The company specializes in industry segments of technical services and technical research, services for hosting companies, hospitals and peer companies, etc., our technical services are of high quality and fast response time!