Working principle of oral comprehensive therapy machine with oral equipment
Turn on the power switch of the air compressor to generate compressed air with a pressure of 0.5mpa~0.7mpa for the machine head, turn on the control switch of the ground box, and turn on the water, gas and power supply. Turn on the power switch of the cold light surgical lamp and it will light up. Pull the arms of the three guns on the dashboard and press the water and air buttons respectively. You can get water jet and air jet. If you press the water and air buttons at the same time, you can get foggy water to meet the different needs of treatment. Pull the arm of the high-speed or low-speed mobile phone on the dashboard and press the foot switch. Compressed air and water reach the nose through the controllers and valves of the air and water systems respectively, driving the turbine to rotate, thus driving the turning needle to rotate, thus achieving the purpose of drilling teeth. When the needle rotates, clean water is sprayed from the nose to reduce the temperature generated when drilling teeth. When the foot switch is released, the nose stops turning, and doctors can choose high-speed or low-speed mobile phones according to the needs of treatment.