The master switch is under the cab or bar, and there is a remote control next to the seat. The remotes are usually controllable, some are buttons.
Before the first use, a syringe needle should be used to put 1 to 2 drops of high-speed lubricant in the shaft. Under normal conditions, the head should be cleaned once a day with cleaning lubricant, and the miniature bearings should be cleaned once after every 2 weeks of work. Normal 0.2 ~ 0.25Mpa working pressure should be maintained; when there is no water, the handset can not be idle, otherwise it will damage the bearings. Needle with blunt to replace the new needle in time, otherwise it will also affect the life of the bearing.
Installation:
The comprehensive dental treatment chair is the necessary medical equipment for dentists, since the world's first electric dental drill in 1887, dental treatment machines in terms of performance and intelligence has improved greatly. In recent years, the application of a large number of new technologies to make dental equipment more humanized. Dental comprehensive treatment machine, divided into pneumatic and electric two kinds.
Gas, water, electricity is a comprehensive dental chair necessary three elements, so before installation must be gas, water pipeline cloth (preferably from the underground), in the water circuit is best to install the initial water filter, so as not to affect the life of the water cartridge in the machine.
The air pump should not be too close to the treatment chair, so as not to affect the patient's treatment by the noise when pumping. If there are more equipment, it is recommended to take the central gas supply, which not only saves money, reduces failure, but also protects the quality of gas supply.