"Hospital disinfection and supply center" "management norms" (WS310.1-2009) stipulates that the washing water should be supplied with tap water, soft water, purified water, etc., and the purified water should comply with the conductivity ≤ 15μs/cm. "Cleaning, disinfection and sterilization technology operation norms" (WS310.2-2009) also requires that: the use of soft water, purified water for the final treatment of instruments, utensils and articles after rinsing, the final rinsing and fine washing should use purified. After rinsing instruments, appliances and articles for final processing, the final rinsing and fine washing to use purified.
Hospital supply room water treatment equipment water production process:
Water intake - raw water buffer tank - - - Multi-media filters -- activated carbon filter -- scale inhibitor -- 5 micron precision filter -- reverse osmosis water purification unit --? Produced water aseptic storage tank -- ultraviolet sterilization device -- point of use -- circulating backwater after ultraviolet sterilization recycling to the water point;
Disinfection supply center to undertake All departments of the hospital all re-use diagnostic and treatment instruments, appliances and articles cleaning and disinfection, sterilization and supply of sterile goods, the final cleaning and sterilization of medical equipment needs to be sufficient supply of purified water, purified water equipment to ensure the safe use of other equipment, to play the maximum benefit of the equipment, the supply of purified water can be called the cornerstone of the normal operation of the disinfection and supply center.