I. LC10Atvp high performance liquid
Detectors SPD a 10Avp (UV-VIS), SPD-M10Avp (PDA)
RF a 10AXL (fluorescence), RID a 10A (refractive index), CDD a 10A (conductivity) ) Autosampler SIL a 10ADvp (high-performance general analysis), SIL a 10AP (preparation, bulk injection), SIL-HT (high efficiency), SIL-10AF (with pre-treatment function)
Column temperature box CTO a 10ACvp, CTO a 10Avp, CT0-10ASvp
Combination of high pressure and low pressure gradient elution. Low-pressure gradient elution. Solution switching and other forms
Data processing CLASS-VP (workstation)
Data processor C-R7A/C-R8A
Pumps LC-IOADvp micro-volume dual-plunger LC-IOATv tandem dual-plunger
Pumps LC-IOADvp micro-volume dual-plunger LC-IOATv tandem dual-plunger
Pumps LC-IOADvp micro-volume dual-plunger < /p>
Constant volume infusion
Flow rate setting range 0.001-5ml/min[1.0-39.2MPa 5.001-9.999ml/min[1.0-19.6MPa]
Flow rate accuracy within ±2% or ±2ul/min(the greater of which)
(O.01-5ml/ min(LC-IOADvp)) min(LC-10ATvp), 0.1-5ml/min(LC-IOATvp), 1.0-39.2MPa, water, room temperature 20℃, except for low-pressure gradient infusion. Except for low-pressure gradient infusion...)
Flow rate accuracy ±O.3% or less (RSD: 0.1%) (0.I-5ml/min, 1.0-39.2MPa, water, room temperature 20℃)
Constant-pressure infusion
Pressure setting range 1.0-39.2MPa[0.1MPa step]
Force setting accuracy ±10% or I. OMPa (the greater of which) or less
Pressure Range High Pressure, Low Pressure
Time Program Flow, Pressure, Event.L00P (Program Repeat).10 files. Total 320 segments Accuracy of pressure display Within ±2% or ±0.5MPa (whichever is greater)
Plunger Cleaning LC-IOADvp Micro Volume Dual Plunger: Connects to an auto-cleaner for automatic cleaning (Patent: N02639130) LC-IOATvp Tandem Dual Plunger: Connects to a syringe or a wash pump for cleaning
Dimensions, Weight W260 x H140 x D420 x H140 x D420mm, 11kg
Operating temperature range 4-35°C
For stable performance of the infusion set, use a degasser
Technical specifications: SPD-10Avp SPD-10AVvp
Light source: 10Avp:D2 lamp, 10AVvp ( D2 lamp, W lamp)
Wavelength range: 10Avp:190-600nm,10AVvp:190-900nm
Spectral bandwidth: 8nm
Wavelength accuracy: ±lnm
Wavelength dependence1: ±0.lnm
Drift: 2 x 10-4AU/h or less (room temperature is certain). ) (10Avp: 250nm, 10AVvp: 250nm,600nm, in-cell air)
Noise level 2: earth O.35 x 10-5AU or less (equivalent response 2sec) earth O.5 x 10-5AU or less (equivalent response Isec) (10Avp: 250nm, 10AVvp: 250nm,600nm. Pool air) [ASTM measurement: ±O.25 x 10-5AU program (time constant 2 sec, 250nm pool air)]
Response: 10-segment switching equivalent to time constants O.05, 0.1, O.5, 1.0, 1.5, 2.O, 3.O, 6.O, 8.O, 10.O sec
Scale: Can be set in steps of 0.0001AUFS. 0.0001AUFS step setting to 0.001-2.56AUFS
Zeroing: automatic zero function, baseline function
Pool (optical path length, capacity pressure): 10mm, 8μL, 12Mpa (about 120Kgf/cm2) (there are also options for the preparation of pools, semi-micro pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools. Inert pool. Temperature control pool)
Dual-wavelength simultaneous monitoring: 10Avp: 190-370nm or 371-600nm
10AVvp: 190-370nm or 371-700nm, any dual-wavelength within the 701-900nm
Ratio Chromatography: Output of a set dual-wavelength ratio
Wavelength Scanning Function: Possibility of The Wavelength step: 1-5nm, 5-step setting (2-5nm when using W lamp), 10nm/sec-5nm/sec, 5-step setting (according to the setting of the wavelength step)
Time program: Any one of the detector body or the system controller can be used, wavelength (2 wavelengths), auto-zero, range, flag indication, response, wavelength scanning, event, and so on. Polarity, ON/OFF, Loop (program loop)
Polarity switching: Able
Outputs: Recorder output: 10mv terminal Integrator output: 5, 1.2.4, 1.25, 2.5AU/V6-segment switching, AU/V6-segment switching, AU/V6-segment switching, AU/V6-segment switching, AU/V6-segment switching, AU/V6-segment switching, AU/V6-segment switching, AU/V6-segment switching.
Dual wavelength mode, ch1 side of the chromatographic axis out.
Dimensions and weight: W260 x H140 x IM20mm, 13kg
Use of ambient temperature range: 4-35 ℃
1: single-wavelength mode when the power supply, indicating that the wavelength change in the importance of.
2:Dual-wavelength mode
Main Features
1 The new generation of highly reliable and efficient LC-VP series improves the reliability and durability of the hardware. Enhanced multi-sample continuous analysis capability and safety measures have been implemented to improve analytical efficiency.
2 Flexibility and ExpandabilityThe LC-VP series consists of a variety of high-performance units. While adding new functions to each unit, compatibility with older models is emphasized. Therefore, the most suitable system can be composed according to the customer's needs (purpose of analysis, budget, and existing model).The LC-VP series is flexible and expandable, ranging from semi-trace analysis and routine analysis to semi-preparation and preparation systems. In terms of data processing, you can flexibly use the new generation of 32bit workstation CLASS-VP6.1 or easy-to-operate data processor.
3 Comprehensive support for GLP / GMP to provide IS09001 quality management required in the proof of adaptability, confirm the validity of the operating guidelines and other documents, as well as the system test, system stability calculation software. LC-VP series in each unit with the new VP function, to achieve high efficiency daily validity of the test. We provide comprehensive support for users, from regular inspections of equipment including older models, to providing effective analysis methods and organizing technical exchange sessions and training courses.
Instrument Introduction
Detailed Explanations
1 The new-generation, highly reliable and efficient LC-VP series has improved hardware reliability and durability. Enhanced multi-sample continuous analysis capability and safety measures have been implemented to improve analytical efficiency.
2 Flexibility and ExpandabilityThe LC-VP series consists of a variety of high-performance units. While adding new functions to each unit, compatibility with older models is emphasized. Therefore, the most suitable system can be composed according to the customer's needs (purpose of analysis, budget, and existing model).The LC-VP series is flexible and expandable, ranging from semi-trace analysis and routine analysis to semi-preparation and preparation systems. In terms of data processing, you can flexibly use the new generation of 32bit workstation CLASS-VP6.1 or easy-to-operate data processor.
3 Comprehensive support for GLP / GMP to provide IS09001 quality management required in the proof of adaptability, confirm the validity of the operating guidelines and other documents, as well as the system test, system stability calculation software. LC-VP series in each unit with the new VP function, to achieve high efficiency daily validity of the test. We provide comprehensive support for users, from regular inspections of equipment, including older models, to the provision of effective analysis methods, and organize technical exchange sessions and training courses.
Infusion pumps LC I10ATvp, LC I10ADvp
Detectors SPD I10Avp(UV-VIS), SPD-M10Avp(PDA)
RF I10AXL(Fluorescence), RID I10A(Refractive index), CDD I 10A (conductivity)
Autosampler SIL I 10ADvp (high-performance general analysis), SIL I 10AP (preparation, bulk injection), SIL-HT (high-efficiency), SIL-10AF (with pre-treatment)
Column temperature box CTO I 10ACvp, CTO-10Avp, CT0-10ASvp
Combination mode High pressure, low pressure gradient elution, low pressure gradient elution, low pressure gradient elution. Low-pressure gradient elution. Solution switching and other forms
Data processing CLASS-VP (workstation)
Data processor C-R7A/C-R8A
Pumps LC-IOADvp micro-volume dual-plunger LC-IOATv tandem dual-plunger
Pumps LC-IOADvp micro-volume dual-plunger LC-IOATv tandem dual-plunger
Pumps LC-IOADvp micro-volume dual-plunger < /p>
Constant volume infusion
Flow rate setting range 0.001-5ml/min[1.0-39.2MPa 5.001-9.999ml/min[1.0-19.6MPa]
Flow rate accuracy within ±2% or ±2ul/min(the greater of which)
(O.01-5ml/ min(LC-IOADvp)) min(LC-10ATvp), 0.1-5ml/min(LC-IOATvp), 1.0-39.2MPa, water, room temperature 20℃, except for low-pressure gradient infusion. Except for low-pressure gradient infusion...)
Flow rate accuracy ±O.3% or less (RSD: 0.1%) (0.I-5ml/min, 1.0-39.2MPa, water, room temperature 20℃)
Constant-pressure infusion
Pressure setting range 1.0-39.2MPa[0.1MPa step]
Force setting accuracy ±10% or I. OMPa (the greater of which) or less
Pressure range High pressure, low pressure
Time program Flow, pressure, event.L00P (program repeat).10 files. Total 320 segments
Accuracy of Pressure Display Within ±2% or ±0.5MPa (whichever is greater)
Plunger Cleaning LC-IOADvp Micro-Volume Dual Plunger: Connects to an auto-cleaner for automatic cleaning (Patent: N02639130) LC-IOATvp Tandem Dual Plunger: Connects to a syringe or a cleaning pump for cleaning
Dimensions. Weight W260 x H140 x D420mm, 11kg
Use temperature range 4-35°C
For stable performance of the infusion set, use a degasser
Technical specifications: SPD-10Avp SPD-10AVvp
Light source: 10Avp:D2 lamp. 10AVvp (D2 lamp, W lamp)
Wavelength range: 10Avp:190-600nm,10AVvp:190-900nm
Spectral bandwidth: 8nm
Wavelength accuracy: ±lnm
Wavelength dependence1: ±0.lnm
Drift: 2 x 10-4AU/h Below (room temperature certain) (10Avp: 250nm, 10AVvp: 250nm, 600nm, cell air)
Noise level 2: Below earth O.35 x 10-5AU (equivalent response 2sec) Below earth O.5 x 10-5AU (equivalent response Isec) (10Avp: 250nm, 10AVvp: 250nm, 600nm, cell air) ASTM Measurement: ±O.25 x 10-5AU program (time constant 2 sec, 250nm pool air)]
Response: 10-segment switching equivalent to time constants O.05, 0.1, O.5, 1.0, 1.5, 2.O, 3.O, 6.O, 8.O, 10.O sec
Scale: Available in steps of 0.0001AUFS. 0.0001AUFS step setting to 0.001-2.56AUFS
Zeroing: automatic zero function, baseline function
Pool (optical path length, capacity pressure): 10mm, 8μL, 12Mpa (about 120Kgf/cm2) (also available as options for the preparation of pools, semi-micro-precipitation pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools, inert pools. Inert pool. Temperature control pool)
Dual-wavelength simultaneous monitoring: 10Avp: 190-370nm or 371-600nm
10AVvp: 190-370nm or 371-700nm, any dual-wavelength within the 701-900nm
Ratio Chromatography: Output of a set dual-wavelength ratio
Wavelength Scanning Function: Possibility of The Wavelength step: 1-5nm, 5-step setting (2-5nm when using W lamp), 10nm/sec-5nm/sec, 5-step setting (according to the setting of the wavelength step)
Time program: Any one of the detector body or the system controller can be used, wavelength (2 wavelengths), auto-zero, range, flag indication, response, wavelength scanning, event, and so on. Polarity, ON/OFF, Loop (program loop)
Polarity switching: Able
Outputs: Recorder output: 10mv terminal Integrator output: 5, 1.2.4, 1.25, 2.5AU/V6-segment switching, AU/V6-segment switching, AU/V6-segment switching, AU/V6-segment switching, AU/V6-segment switching, AU/V6-segment switching, AU/V6-segment switching, AU/V6-segment switching.
Dual-wavelength mode, ch1 side of the chromatographic axis output.
Dimensions and weight: W260 x H140 x IM20mm, 13kg
Use of ambient temperature range: 4-35 ℃
1: single-wavelength mode when the power supply, indicating that the wavelength change in the importance of.
2: dual-wavelength mode
Two, LC20A original liquid chromatograph
Shimadzu LC-20A series of liquid chromatograph is Shimadzu developed one of the world's most stable liquid chromatograph models, infusion unit Prominence LC-20AD/20AT/20AB, flexible configuration for different needs of the customer. Function and stability are better than LC-10A and 15C.
LC-20AD
Infusion method: parallel dual plunger
Plunger capacity: 10ul
Maximum discharge pressure: 40MPa
Flow rate setting range: 0.0001mL/min-10.0000mL/min
< p>Flow rate accuracy: within 1% or 0.5ul /min of the greater value (0.01mL/min-2mL/min)Flow rate precision: 0.3 or less (RSD: 0.1% or less)
LC-20AT
Infusion: tandem double-plunger
Plunger capacity: 47ul for the main pump head. 23ul for the secondary pump head
Plunger capacity: 47ul for the main pump head. 23ul for the secondary pump head, Sub-pump head 23ul
Maximum discharge pressure: 40MPa
Flow rate setting range: 0.001mL/min-10.000mL/min
Flow rate accuracy: 2% or 2uL/min within the greater of (0.01mL/min-5mL/min)
Flow rate precision: 0.3% or less (RSD: 0.1%)
Flow rate precision: 0.3% or less (RSD: 0.1%)
Flow rate precision: 0.3% or less (RSD: 0.1%)
Flow rate precision: 0.3% or less (RSD: 0.1%) (RSD: 0.1% or less)
LC-20AB
Infusion method: parallel double plunger (2 types)
Plunger capacity: 10uL
Maximum discharge pressure: 40MPa
Flow rate setting range: 0.0001mL/min-10.0000mL/min
Flow rate accuracy: within 1% or 0.5uL/min of the greater value (0.01mL/min-2mL/min)
Flow rate precision: 0.3 or less (RSD: 0.1% or less)
Gradient system
LC-20AB
Gradient mode: high-pressure mixing
Number of solvents mixed: 2 liquids
Mixing concentration accuracy: below 0.5%
LC-20AD/20AT high pressure GE specifications
Gradient method: high pressure mixing
Number of mixing solvents: 2 liquids or 3 liquids
Mixing concentration accuracy: below 0.5% (LC-20AD), below 1% (LC-20AT)
LC-20AD/20AT low-pressure GE specifications
Gradient method: low-pressure mixing
Number of solvents to be mixed: up to 4 liquids
Mixing concentration accuracy: 0.5% or less (LC-20AD), 1% or less (LC-20AT)
Ultraviolet visible detector
Prominence SPD-20A/20AV UV-VIS Detector
Prominence SPD-M20A PDA Detector
SPD-20A / SPD-20AV
Light Source: SPD-20A D2 Lamp, SPD-20AV D2 Lamp, W Lamp
Number of Diode Elements: none
< p>Wavelength range: SPD-20A 190nm-700nm , SPD-20AV 190nm-900nmSpectral bandwidth? Gap width: 8nm
Wavelength accuracy: below 1nm
Wavelength precision: below 0.1nm
Noise: ±0.25×10-5AU (under specified conditions)
Drift: 1×10-4AU/h (under specified conditions)
Linearity: 2.5AU (ASTM specification)
Function: 190nm-370nm or 371nm or more dual-wavelength detection, proportional chromatogram output, wavelength scanning
Cell: Optical path length: 10mm, Capacity: 12mL, Pressure-resistant: 12MPa
SPD-M20A
Light source: D2 lamps, W lamps
Number of diode elements: 512
Wavelength range: 190nm-800nm
Spectral bandwidth? Slit width: 1.2nm (high resolution mode) 8nm (high sensitivity mode)
Wavelength accuracy: below 1nm
Wavelength precision: below 0.1nm
Noise: ±0.3×10-5AU (under specified conditions)
Drift: 0.5×10-3AU/h (under specified conditions)
Linearity: 2.0AU (ASTM specification)
Functions: Contour Output, Spectral Retrieval, Maximum Output
Cell: Optical Path Length: 10mm, Volume: 10mL, Pressure Resistance: 12MPa
Autosampler Prominence SIL-20A/20AC
Injection Mode: Full Volume Injections, Inlet Volume Variable
Sample volume setting range: 0.1ul-100ul (standard), 1ul -2.000mL (optional)
Sample handling capacity:
175 (1mL vial)
105 (1.5mL vial) (SIL-20AC 70 (1.5mL vial))
50 (4mL vial)
192 (2 pieces of 96-well MTP)
768 (2 pieces of 384-well MTP)
Ten 1.5mL vials can be used in addition to each of the above sizes
Injecting Volume Accuracy: 1% or less
Injecting Volume Precision: RSD: 0.3% or less (under specified conditions at 10mL injection)
Cross-contamination: 0.005% or less (naphthalene, chlorhexidine)
Repeated injection times: Maximum 30/1 sample
Needle cleaning: set arbitrarily before and after the injection
Sample cooler: SIL-20A No SIL-20AC Module cooling heating method, and moisture removal Function 4°C-4
Sample cooler: SIL-20A No