How to proofread electronic scales accurately

The accurate calibration methods of electronic scales include: detection before debugging the accuracy of electronic scales, weight measurement, calibration of four corners and center points, confirmation that the empty scale is "0" and repeatability error calibration.

1, precision test of electronic scale before debugging

Check whether the appearance of the electronic scale is clean and the switch is in good condition, then check whether the power supply is sufficient, whether the digital display on the instrument is bright, complete and clear, and whether the zero position of the electronic scale is correct.

2. Weight measurement

Electronic scales generally have weights, so it is more accurate to use these weights. Measure with standard weights of different masses. The weight of the weight should be within the range of the electronic scale. Measure the weight in turn. During this process, observe whether the number displayed on the display is consistent with the weight of the weight. Try it several times, find out the mistakes, and then make adjustments.

3, four corners and center point calibration

After the electronic scale is horizontally adjusted, it should be as close as possible to the weighing platform of the full scale to stabilize all parts of the weighing platform. Then add a certain weight to the scale for calibration. After the initial calibration, remove the weights and adjust the four corners. Press the angle with a proper weight and check the four-corner error. The general basic error is within1d. If it is smaller, it should be realized by adjusting the value of angle coefficient K.

4. Confirm that the empty scale is "0"

After debugging the load-bearing key, evenly add the weight equivalent to the maximum weight to the load-bearing platform. After the display value is consistent with the weight value according to the position and method introduced in the instructions of various weighing displays, remove the weight and confirm that the empty scale shows "0".

5, repeatability error calibration

Calibration of repeatability error of electronic platform scale: Take the maximum load (full scale) weight of electronic platform scale, put it in the center of electronic platform scale for four times, and record the reading after each stabilization.