How should I maintain my Medtronic power system?

Hello, I am Hangzhou Si-Han Medical Equipment Co. We are a professional endoscopic system after-sales service provider.

Provide power system maintenance such as: Medtronic, Stryker, Schlafly Smith&Nephew, Dyonics Dianli, Storz Storz, Wolf Wolf, Linvatec Linvatec, Aesculap Snake, Bienair, German Medicon, Johnson & Johnson Anspach, Synthes Synthes, and so on. Medicon, Johnson Anspach, Synthes, RUDOLF, Ackerman, and so on.

On how to maintain the power system, for you to answer the following:

(A) cleaning

1, the initial wash. After the completion of surgery, the use of running water to rinse the surface of the instrument blood stains, open the elbow joint parts, clamping parts with a brush to scrub, and then the use of high-pressure water gun will be the joints, lumen and other parts of the blood stains for repeated rinsing.

2, enzyme washing. After the initial washing of the instruments into the multi-enzyme solution soaked for 5min, removed and washed with running water, and then put it into the multi-enzyme ultrasonic instrument cleaner for 8min, clear the various parts of the protein, blood stains, fat and other organic matter is not easy to clean, washed with running water and then repeatedly scrubbed with a brush. Then placed into the acidified water immersion for 2min, followed by distilled water to wash out the lumen parts with a high-pressure water gun rinse, placed into the medical lubricant soak for 1min, followed by absorbent soft cloth to dry, and high-pressure air gun to blow dry.

(2) Maintenance

1, after the operation to turn off the instrument, cut off the power supply, to avoid sudden power and power failure caused by instrument damage. Positioning storage, plus dust cover, reduce the number of moving to avoid damage.

2. Mirrors and other mirrors should be gently wiped with dustproof paper to avoid abrasion and collision; instruments should be gently held and placed to avoid falling or loss.

3, a monthly rust and lubrication maintenance, after completion of the use of special lubricants sprayed again on the surface of the instrument.

4, found damaged or screws loose, etc. to deal with, to ensure that the surgery required.

Hope to help you, the following is the power host of common failures and solutions, as well as our maintenance of the power host of the case

Power host of common failures and solutions:

1. power board is bad: repair or replacement of the power supply board

2. bad motherboard, motor short-circuit, drive circuit burned out: replacement of the motherboard

3. Sensor bad: repair or replace the sensor

4. Display bad: replace the display

5. Display board bad: repair or replace the display

Maintenance cases:

Brand: Medtronic Medtronic

Model: power host EC300

1. Appearance inspection: check the appearance of the host intact, no obvious signs of external damage.

2. Failure phenomenon: the host is connected to the AC power, the boot directly reported errors, code:34, code:27, code:26. can not be used.

3. dismantling diagnosis: replace the board test, replace the good driver board found that the fault disappeared, judgment for the driver board has a circuit failure, resulting in the output circuit and thus no output, and there are a number of chip burned short circuit, the board signal error.

4. Maintenance program: ① replace the handle driver board ② calibrate the output voltage, reinforce the internal firmware.

5.