1 Before working daily, use a dry cloth to clean the gastrointestinal machine 1 time, and form a work system.
2 Always pay attention to the position of each regulator on the console surface to see if there is any looseness or displacement.
3 During exposure, you should pay close attention to the indication and display of the gastrointestinal machine's instrumentation, and pay attention to any abnormal sounds.
4 The bed surface of the examination bed should be kept clean, and if there is patient's vomit, feces or blood, it should be removed in time so as not to cause harm to the gastrointestinal machine.
5 Always pay attention to the power supply voltmeter, kilovoltmeter, milliampere meter and other kinds of instructions, display meter parameters are normal, if there is a high or low, trembling, rapid impulse, zero drift and other phenomena, should be immediately shut down, and report to the maintenance personnel.
6 fluorescent screen, monitor, display and other components when not in use, should be covered by a cloth cover to avoid sunlight or light irradiation, aging phenomenon. If the fluorescent screen aging caused by the decline in brightness, should be replaced in a timely manner with a new fluorescent screen.
7 half a day a week to maintain, maintain the machine, including gastrointestinal machine cleaning, mechanical parts of the lubrication, circuit check and necessary adjustments, system reliability and stability check.
8 Pay attention to the observation of the sensitivity of the activities of the various moving parts of the gastrointestinal machine, there can not be too much friction, and often in the bearings, the track coated with lubricant, to increase its flexibility, to reduce mechanical wear and tear.
9 parts of the wire rope should always check whether there is wear and tear of the "broken" phenomenon, if any should be updated in a timely manner.
10 to often check the gastrointestinal machine between the parts of the fixed screws, nuts, pins have no loose phenomenon, found that the problem should be dealt with in a timely manner, we must often check the main components of the protection of the ground is good, to prevent the loosening of the situation, rust and corrosion.
11 establish gastrointestinal machine machine files. Gastrointestinal machine should be stopped immediately when a malfunction occurs, and report to the superior and maintenance personnel. Promptly record the circumstances surrounding the malfunction and the repair.
12 Quality control and maintenance of gastrointestinal machine is carried out by the responsible engineer, who regularly inspects the wearable parts and reports to the head of the department on a regular basis. Or contact the medical equipment company manufacturer.