A. Basic parameters of intracavity electrode:
1. Pulse frequency: 1000Hz±150Hz;
2. Pulse voltage (peak-to-peak): 100V±20V;
3. Maximum temperature of the working surface: 38℃~40℃;
4. Dimensions of working part of intracavity treatment head: φ25mm×150mm
5. Voltage adjustment mode: 20 levels, graded and adjustable;
6. Temperature adjustment mode: 16 levels, graded and adjustable from the ambient temperature to the maximum temperature of the working surface.
B. Basic parameters of abdominal treatment
1. Pulse frequency: 1000Hz±150Hz;
2. Pulse voltage (peak value): 100V±20V;
3. Maximum temperature of the working surface: 42℃~45℃;
4. Temperature adjustment mode: 16 levels, adjustable in grading from the ambient temperature to the maximum working surface temperature. temperature, graded adjustable.
5, abdominal treatment tablets treatment area: 381.5c ㎡
C, other:
1, microprocessor LCD performance should be stable, in the room under the natural light, 30cm away from the display, the display should be clearly visible.
2. The connecting plugs of intracavitary treatment head and abdominal treatment pad should be with locking device. Product Classification: Class II Medical Devices Anti-Shock Type: Class I Device Anti-Shock Degree: BF Type Device
Product Characteristics: Mobile Device
Rated Power: 400VA Fuse Capacity: 2A
Medical Devices Product Registration Certificate No.: E Food & Drug Supervision and Inspection of Machinery (Quasi) Zi No. 2250951 of 2006
Product Standard Number: YZB/Er 0497-2006 Ambient temperature: 5℃~40℃;
Relative humidity: not more than 80% RH;
AC power supply: 220V±22V, 50Hz±1Hz.