What are the brands of imported and domestic anesthesia machines? full marks

Product name: ry-ii b+ anesthesia system

Model: ry-iib+

Status: Normal

Brand:

Origin: Jiangsu

Quotation:

Number of visits: 166

Main parameter

Flow range: (0.25 ~15) l/min; Temperature range: (15 ~ 35)℃;

Product characteristics

RY-IIB+ anesthesia system is a new type of anesthesia system with modern microcomputer control technology, which is suitable for modern anesthesia surgery. The appearance is novel, the connection is convenient, the performance is excellent, and the price is favorable, which constitutes the best cost performance and allows users to enjoy great value. Ry-IIB+ anesthesia system has strong practicability, and has been well tested and tested in clinic, which has won unanimous praise from customers.

Product parameter

Anesthesia machine

1. Working mode: fully closed, semi-closed (0.5 ~ 7.5) kPa, semi-open;

2. Gas supply: central pipeline or gas cylinder can be used for gas supply, with standby gas source input port, with rated working pressure of 0.4 MPa and allowable fluctuation range of (0.27 ~ 0.55)MPa;;

3. Flowmeter: O2: (0.1~10) l/min; N2O:(0. 1~ 10) l/min;

4. Rapid oxygen supply flow: (35 ~ 75) l/min;

5. High-precision evaporator: equipped with KT-III enflurane and isoflurane evaporators.

Concentration range: 0 ~ 5 vol %;;

Flow range: (0.25 ~15) l/min;

Temperature range: (15 ~ 35)℃;

Automatic compensation of flow, temperature and pressure.

Anesthetic ventilator

1, inspiratory flow: (6 ~ 66) l/min;

2. Adjustable range of tidal volume: (50 ~1500) ml;

3. Frequency: (6 ~ 99) times/minute;

4. Breathing-breathing ratio: 3 ∶ 1 ~ 1 ∶ 9 (inspiratory time 0.3s ~ 6.0s);

5. Upper pressure alarm: (1.0 ~ 6.0) kPa (not lower than the lower limit);

6. Lower pressure alarm: (0.4 ~ 4.0) kPa (delay 4 s ~15 s);

7. Maximum safety pressure: the exhaust pressure of the pressure limiting valve is (5.1~ 6.0) kPa;

8. Low ventilation alarm: when the exhaled tidal volume is lower than the minimum set value, an alarm should be given within 90 seconds;

9. The tidal volume range of bellows: adult: (300 ~1500) ml; Children: (50 ~ 300) ml;

10, power failure alarm: when the power supply is interrupted, an audible alarm will be given, and the alarm duration shall not be less than120s; ;

1 1. gas supply interruption alarm: low pressure alarm will be triggered when gas supply is interrupted;

12, alarm mute time: no more than 120s.

Product name: GSM-Ⅲ (c) anesthesia machine

Model: GSM-Ⅲ (C)

Status: Normal

Brand:

Origin:

Quotation:

Number of visits: 3 17

Product parameter

▲ Anesthesia host part:

● 1. Mode: closed, semi-closed and semi-open.

●2. Gas supply source: O2 and N2O pressure is 0.3 ~ 0.5 MPa.

●3. Flowmeter: O2 and N2O two-tube flowmeters 0 ~ 10L are equipped with linkage mechanism to ensure that the oxygen concentration is not lower than 25%.

●4. Rapid oxygen supply: 35l/min ~ 75l/min.

5. Electronically controlled pneumatic constant volume anesthesia ventilator

●6. Low air source alarm: When the O2 air source is too low, there will be an audible alarm for more than 7 seconds.

●7. Upper pressure limit: When the airway pressure exceeds the upper limit set value, the breathing opportunity will be automatically limited to ensure that the airway pressure can not continue to increase and ensure safety.

●8. Double-end differential pressure flow sensor can monitor tidal volume, minute ventilation, respiratory frequency, airway pressure, pressure waveform, flow velocity waveform (bidirectional), upper and lower pressure limits, asphyxia and oxygen concentration in real time in both mechanical and manual modes.

9. The external suspension system is exquisitely designed and easy to operate.

▲ anesthetic evaporation part:

● High-precision British imports (Penlon cans), double cans (enflurane, isoflurane)

● Evaporator has the function of automatic compensation for flow, pressure and temperature.

● concentration adjustment range: 0.2 ~ 5% vol

● Evaporator has self-locking and interlocking functions.

▲ Anesthesia ventilator part:

● 1. Scope of application: adults and children.

●2. Drive: pneumatic electric control.

●3. Display: 10.4 color TFT large screen display is adopted.

●4. Mode: Manual and Automatic.

●5. Tidal volume: 100 ~ 1500ml.

●6. Frequency: 6 ~ 60 BPM

●7. Breathing ratio: adjustable from 3: 1 to 1: 4.

●8. Upper pressure limit: 2 ~ 6 kPa

●9. Lower pressure limit: 0.6 ~ 2 kPa

● 10. Monitoring parameters: tidal volume, respiratory frequency, respiratory ratio and peak airway pressure.

Product introduction

▲ Gift configuration:

● Children's bellows: 50 ~ 300ml

● Pediatric circuit

▲ Optional:

● Optional end-expiratory carbon dioxide monitoring.

● An external printer can be selected to record alarm events.

Product name: anesthesia machine PN-2000M

Model: PN-2000M

Status: Normal

Brand:

Origin:

Quotation:

Number of visits: 322

Product parameter

Technical parameters of anesthesia machine

Flow adjustment range:

N2O:0. 1L/min ~ 12L/min,

Oxygen: 0. 1 l/min ~ 1.2 l/min

Rapid oxygen supply flow: 35l/min ~ 75l/min.

Oxygen supply low pressure alarm, N2O automatically cut off:

When the oxygen supply pressure is lower than 0.25Mpa, an audible alarm will be given, and the duration will not be less than 7 seconds.

When the oxygen supply pressure is lower than 0. 1Mpa, the N2O flow is automatically cut off.

An and isoflurane evaporators (imported or domestic):

Concentration range: 0.25% ~ 5% by volume

Flow range: 0.25l/min ~ 15l/min.

Temperature range:15℃ ~ 35℃ Automatic compensation of flow, temperature and pressure.

Anesthetic ventilator

Tidal volume: 200ml ~ 1200ml.

Respiratory frequency: 10 beats/min ~ 30 beats/min

Respiratory ratio: 1: 1 ~ 1: 3.

Upper pressure limit (limit): 1 kPa ~ 6 kPa

Lower pressure limit: 0 ~ 4 kPa

Bellows: 100ml ~ 1500ml

Product introduction

Product description:

PN-2000M anesthesia machine

Equipped with high-precision enflurane and isoflurane evaporators, high-precision N2O and O2 flowmeters, emergency oxygen supply device, oxygen supply low-pressure alarm and N2O automatic cut-off. The patient circuit system conforms to ISO standard and has the functions of manual breathing and mechanical ventilation, which makes anesthesia safer and more convenient.

PN-2000MII anesthesia ventilator

Microcomputer-controlled pneumatic electronic control ventilator has intuitive operation and complete functions.

Product name: multifunctional anesthesia machine PN-200OMD

Model: PN-2000 OMD

Status: Normal

Brand:

Origin:

Quotation:

Number of visits: 349

Product parameter

Technical parameters of anesthesia machine

Flow adjustment range:

N2O:0. 1L/min ~ 12L/min,

Oxygen: 0. 1 l/min ~ 1.2 l/min

Rapid oxygen supply flow: 35l/min ~ 75l/min.

Oxygen supply low pressure alarm, N2O automatically cut off:

When the oxygen supply pressure is lower than 0.25Mpa, an audible alarm will be given, and the duration will not be less than 7 seconds.

When the oxygen supply pressure is lower than 0. 1Mpa, the N2O flow is automatically cut off.

An and isoflurane evaporators (imported or domestic):

Concentration range: 0.25% ~ 5% by volume

Flow range: 0.25l/min ~ 15l/min.

Temperature range:15℃ ~ 35℃ Automatic compensation of flow, temperature and pressure.

Anesthetic ventilator:

Tidal volume: 200ml ~ 1200ml.

Respiratory frequency: 10 beats/min ~ 30 beats/min

Respiratory ratio: 1: 1 ~ 1: 3.

Upper pressure limit (limit): 1 kPa ~ 6 kPa

Lower pressure limit: 0 ~ 4 kPa

Bellows: 100ml ~ 1200ml

Product introduction

Product description:

PN-2000MD multifunctional anesthesia machine

It consists of anesthesia machine and anesthesia ventilator. The anesthesia machine microcomputer controls the pneumatic electronic control ventilator, which has intuitive operation and complete functions. Anesthesia ventilator is equipped with high-precision enflurane and isoflurane evaporators, high-precision N2O and O2 flowmeters, emergency oxygen supply device, oxygen supply low-pressure alarm and N2O automatic cut-off. The patient circuit system conforms to ISO standard and has the functions of manual breathing and mechanical ventilation, which makes anesthesia safer and more convenient.

Product name: ry-Ⅱ a anesthesia machine

Model: ry-IIA

Status: Normal

Brand:

Origin: Jiangsu

Quotation:

Number of visits: 153

Main parameter

Concentration range: 0 ~ 5 vol %;; Temperature range: (15 ~ 35)℃; Flow range: (0.25 ~15) l/min;

Product parameter

Gas supply pressure: the rated working pressure is 0.4MPa, and the working pressure range is (0.27 ~ 0.55) MPa, and the gas is supplied by gas cylinders or central pipelines;

Flow rate: O2: (0. 1 ~ 10) L/min, N2O: (0.1~10) l/min;

Flow control valve: the linkage mechanism of oxygen and nitrous oxide ensures that the oxygen concentration is not lower than 25%, and nitrous oxide cannot enter the anesthesia respiratory system when the oxygen flow control valve is not opened;

Rapid oxygen supply flow: (35 ~ 75) l/min;

Respiratory circuit: working mode: fully closed, half closed and half open (1~ 4) kpa; Exhale and inspiratory impedance: no more than 0.6kPa;;

Evaporator: equipped with KT-II enflurane or isoflurane evaporator.

Concentration range: 0 ~ 5 vol %;;

Temperature range: (15 ~ 35)℃;

Flow range: (0.25 ~15) l/min;

Automatic compensation of flow, temperature and pressure.

Product introduction

RY-IIA anesthesia machine can meet the basic requirements of clinical use. The machine has reasonable layout and simple structure, and is suitable for hospitals at all levels. The loop system conforming to ISO standard has various working modes, such as fully closed, semi-closed and semi-open. It is suitable for children and adults over 4 years old, and the cases are not limited. When used for children under 4 years old, the pediatric anesthesia circuit can be purchased from the manufacturer.

Product name: American VME animal anesthesia machine

Model: VME

Status: Normal

Brand:

Origin: USA

Quotation: negotiable

Number of visits: 193

Product characteristics

1, professional closed respiratory circulation loop:

2. The oxygen flowmeter is accurate, the output is stable, and it has the function of rapid oxygenation. It has the function of rapid oxygenation; You can choose to equip with a double flow meter.

3. It has the safety function of APL automatic pressure relief: to avoid pressure injury to animals caused by excessive air pressure.

4, with pressure detection function

5. Patented air inlet valve: allows animals to breathe the atmosphere after the oxygen supply stops. Ensure the safety of animals in unexpected situations.

6, with carbon dioxide absorption system:

7, with exhalation and inhalation valves: can clearly observe the breathing state of anesthetized animals: inhalation or exhalation.

8. Compact structural design, flexible movement, light handle and convenient use;

Product Name: TIAC-I Intravenous Anesthesia Controller (Target Controlled Pump)

Model: TIAC-I

Status: Normal

Brand:

Origin: Guangdong

Quotation:

Number of visits: 199

Main parameter

Precise control: the capacity error is 2%

Product characteristics

1, through the clinical study of 18, it is the only intravenous anesthesia control device administered according to the pharmacokinetic and pharmacodynamic parameters of Chinese people, and the anesthesia control is stable;

2. Seven kinds of intravenous anesthetics commonly used in clinic were selected, and the data were all from clinic.

3. First-class imported injection system is selected, with stable performance;

4, large LCD screen, Chinese menu interface, simple operation;

5. There are two management modes: target control mode and constant speed mode. In the target control mode, plasma or effect room concentration mode can be selected for administration.

6. Own independent intellectual property rights, obtain 3 national patents, publish more than 90 academic papers and 2 books;

Product parameter

Accurate control: the capacity error is 2%;

After the battery is fully charged, it can work continuously for more than 6 hours.

Product introduction

At first, an "intelligent intravenous anesthesia controller" with independent intellectual property rights was developed by using China intravenous anesthesia parameters and world-class hardware.

Personally, according to my own experience and understanding, I think the machines in Yi 'an, Beijing are better.

Not to belittle domestic manufacturers, domestic anesthesia machines are much worse than imported machines in terms of stability, ease of use and safety. If you remember a key part when you choose, as long as it is imported, the performance of this anesthesia machine is mostly good, and it is not recommended to make it purely in China.

Doelger from Germany and Omida from America are recommended. Don't worry.