Ultrasound machine maintenance should pay attention to what aspects?

1 .B ultrasound machine is a precision electronic instruments, there should be a person responsible for the daily cleaning and wiping work, especially on the probe

head of the daily maintenance is particularly important, after the use of the probe must be wiped clean, because the probe above the residual coupling agent is corrosive

sexual, if the maintenance is not good will shorten the service life of the probe. Probe cable due to repeated twisting prone to cause breakage or contact

bad, use should be careful not to pull or excessive twisting.

2 to pay attention to the ultrasound machine cooling and ventilation, once a month to check and clean the machine ventilation filter. Ensure

machine work without overheating and burn the machine.

3 toilet paper used to choose a better quality, because the long-term use of poor quality toilet paper, paper impurities and dust in the paper is too much, a serious impact on the cleanliness of the machine room, more importantly, these impurities and dust will be sucked into the machine inside, resulting in the machine

Can not be well ventilated heat dissipation, long time working machine is easy to burn components.

4 to strengthen regular maintenance and the use of personnel to strictly comply with the operating procedures is to reduce the failure to extend the life of the machine is an important means of cleaning, generally half a year, the operating time to maintain good ventilation, to prevent overheating, overheating, continuous power-on time is not more than 4h, the mechanical fan sweeping probe for each continuous use of l0min 15min to stop sweeping once and to have a good regulated power and reliable grounding, etc. These are the main reasons why we are so interested in the machine. Grounding and so on, these are we must strictly abide by, so that can greatly reduce the maintenance rate of ultrasound machine, improve the instrument

The use of equipment.

5 ultrasound visualizer, especially the more sophisticated ultrasound visualizer requires a certain working temperature conditions, the optimal temperature of about 25 degrees Celsius, the highest not more than 35 degrees Celsius, the minimum can not be lower than 10 degrees Celsius. In special areas should be configured with air-conditioning equipment, in

Use of air-conditioning equipment, it should be noted that often due to the original high relative temperature, when the temperature plummeted to zero, the imaging instrument within the

components and terminal boards on the condensation of water droplets and energized, thus damaging the electronic components, so to slow down the cooling process, and to make air conditioners to keep

Holding a continuous work, the best is the first moisture absorption and dehumidification, the best is to reduce the relative humidity before turning on the ultrasound equipment. Reduce the indoor relative humidity before turning on the air conditioning unit.

6 to prevent bright light exposure, such as: sunlight, fluorescent lamps or other bright light illumination are prone to oscilloscope screen phosphor aging,

luminous rate of dimming, other fixtures: such as the probe line, camera lenses should not be exposed to sunlight, should be the windows of the room should be

good sunshade, every time after checking the whole machine with a shading cloth shading.

7 continuous working time of the instrument: in the unit of time, the power should not be too frequent, so as not to shorten the oscilloscope tube and the instrument life

life, in the temperature of the appropriate environment, generally can be three to four hours of continuous work, such as no air-conditioning equipment, the general work of the left two hours

right, turn off the machine for 20 to 30 minutes in order to facilitate the dissipation of heat, the temperature is cooled down. Because of the daily work of the best will be focused on the patient together with a fixed

Time check to improve the service life of the instrument per unit of time.