How to calibrate Mindray 220?

The calibration of BS-220 biochemical instrument you mentioned is also called calibration. Please describe the problem clearly in the future.

First of all, you need to prepare the calibration and quality control materials for Mindray.

1, in the "calibration"-"calibration solution settings" including calibration solution name, batch number, period of validity; The location of the calibration solution, the calibration items of the calibration solution and the corresponding standard concentration.

2. Select the calibration rules of each item in "Parameters"-"Calibration Rules". Generally, it is a two-point linear calibration, and you need to provide two concentrations of standards, one is Mindray's supporting calibrator, and the other is water with a concentration of 0. Water should also be set in the calibration solution setting first, and two standards must be checked in the calibration rules interface for each item, otherwise it cannot be calibrated.

3. After completing the above two steps, you can apply for calibration in the "Calibration"-"Calibration Application" interface, where the gray items are those without calibration fluid information or calibration rules, and the white items are those that can apply for calibration.

4. After the calibration is completed, do quality control to confirm whether the result is reliable.