Requirements for equivalent thickness of shielding protection line in different types of X-ray equipment rooms;
1, and in a photographic room with a nominal value of 125kV or more, the lead equivalent mm in the useful beam direction and the lead equivalent mm in the useful beam direction are 3: 2.
2. The lead equivalent mm in the useful beam direction and the lead equivalent mm in the useless beam direction of the photographic room, oral CT room and dental panoramic room (with skull photography) with a nominal value of 125kV or less are 2: 1 respectively.
3. The lead equivalent mm in the useful beam direction and the lead equivalent mm in the useless beam direction in the fluoroscopy room, bone densitometer room, dental dental dental film room, dental panoramic room (without skull photography) and breast room are 1: 1 respectively.