The more challenging the cleaning process, the larger the sample size for the test. Typically, a validation test requires three, six, or nine instruments, depending on the complexity of the instrument structure.
The sample volume under test is considered sufficient if the following three conditions are met:
(1) There are at least 3 data for each analyte to be tested
(2) All data are within acceptable indicators of cleaning effectiveness
(3) The data with the largest value plus the standard deviation of the data does not exceed the acceptable indicators of cleaning effectiveness
At the same time, it is important to Consider the positive and negative control samples used during the test. We recommend going to Tacteasy, a medical device CRO, which is more efficient.