Operating room it power supply 8kva which specification has requirements

Medical IT isolated power supply system for the operating room, should meet the following requirements, and relevant national standards:

1, isolated power supply system to meet the GB 16895.24-2005/IEC60364-7-710 standard.

2, medical isolation transformer meets IEC61558-2-15 standard.

3, insulation monitoring alarm system in line with IEC61557-8 standard.

4, real-time monitoring of the isolation power supply system to ground insulation condition, according to the need to set different alarm response value.

5, real-time monitoring of the isolation transformer load conditions and transformer winding resistance temperature

6, real-time monitoring of the isolation power system components and the system grounding wiring conditions.

7, real-time display of isolation power system load percentage, system insulation resistance.

8. Adaptive pulse signal measurement method detection, in order to improve the anti-interference ability of the system detection.

1 isolation transformer

1.1 to meet the "building electrical installations Part 7-710: special installations or places to get the requirements of the medical premises" GB16895.24-2005 / IEC60364-7-710: 2002 710.512.1.6 in the provisions of the medical isolation transformer should be in line with the following IEC61558-2-15 requirements.

1.2 The rated voltage Un on the secondary side of the transformer should not exceed 250 V. The medical IT system for mobile and fixed equipment should use a single-phase transformer, whose rated output capacity should not be less than 0.5 KVA but should not exceed 10 KVA.

1.3 The metal case of the medical isolation transformer needs to be coupled with local equipotential coupling terminal boards.