Product Name (Chinese) Whole-body X-ray Computed Tomography Scanner System
Product Name (English) Computed Tomography Scanner System
2, Whole-body X-ray Computed Tomography Scanner System - Specification
< p>LightSpeed Pro163, whole-body X-ray computed tomography scanner system - product standards
YZB/USA 2345 "LightSpeed Pro16 Whole-body X-ray Computed Tomography Scanner System"
4, whole-body X-ray computed tomography scanner system - structure
Structure: by the rack, patient examination bed, console, power distribution, workstation, bulb tube and camera composition. Performance: rack opening diameter of 70cm; spatial resolution of 10% MTF contrast case for 13Lp/cm; low-contrast resolution (20cm CATPHAN model) in the case of 0.3% contrast in 13.3mGy 5mm, in 37.2mGy 3mm; reconstruction time of less than 1.5s; the minimum thickness of the body layer is * 0.625mm; Scan frame tilt angle range: front/back ±30 degrees.
5, whole-body X-ray computed tomography system - Scope of application
The product is suitable for whole-body computed X-ray tomography examination.