Cleaning, drying to be sterilized treatment of items immersed in 2% glutaraldehyde solution, cover, soak 10h, aseptic operation to remove, rinse with sterilized water, and aseptic procedures to dry and ready for use, alkaline glutaraldehyde use 14 days.?
Disinfection treatment: immersion method. Will be sterilized treatment items immersed in 2% glutaraldehyde solution, cover. For bacterial propagule contamination, soak for 10min, hepatitis virus contamination soak for 30min, remove and rinse with sterilized distilled water and wipe dry.
Wipe method: wipe the surface contaminated by bacterial propagules with 2% glutaraldehyde solution, disinfecting effect 10min, hepatitis virus contaminated surface, disinfecting effect 30min.
Expanded information
1. Glutaraldehyde is irritating to the skin and mucous membranes, and it is toxic to human beings, and the use of glutaraldehyde using the liquid has a Serious injury. It should be prepared and used in a well-ventilated place, pay attention to personal protection, wear protective mask, protective gloves and protective glasses. In case of inadvertent contact, it should be immediately flushed continuously with water, and early medical attention should be sought if the eyes are injured.
2, should be used in a well-ventilated place, if necessary, the use of places should be exhaust equipment. If the concentration of glutaraldehyde in the air at the place of use is too high, it is recommended to be equipped with self-contained breathing apparatus (positive-pressure protective mask). Containers used for soaking instruments must be clean, covered, and sterilized before use.
3, at room temperature, after adding sodium nitrite and sodium bicarbonate glutaraldehyde disinfectant can be used continuously for a maximum of 14 d. Continuous use of the process, the daily monitoring should be strengthened to grasp the concentration of changes below the required concentration, stop using. After disinfection or sterilization of medical devices, before use in a sterile manner to remove, with sterile distilled water repeatedly rinsed clean, and then dry with sterile gauze, etc. before use.
4, with endoscopic cleaning and disinfecting machine disinfection treatment, the endoscopic cleaning and disinfecting machine used must be approved by the Ministry of Health Health Health Permit, the disinfection procedures used must also be approved in its approval of the use of procedures. The product should be sealed, protected from light and stored in a cool, dry and ventilated place. It shall not be stored in the open air, and shall not be mixed with other toxic substances. It should not be inverted during transportation, and should be protected from pressure, collision, extrusion, sunlight and rain, and the vehicle should be kept dry frequently.